Medical Device Post-Market Surveillance and PSUR Planning

Oct 12
7 min read

Updated: 4 days ago

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 10, 2026.


How should PMS data support an applicable PSUR?


Choose PMS documentation to organize continuing collection and analysis; choose PSUR documentation for the periodic output when it applies. Under EU MDR, a PMS report and a PSUR are distinct outputs with applicability tied to classification and the legal framework. Confirm the device route before choosing the report structure.


Post-market surveillance turns information from device use into decisions about safety, performance and the quality system. A periodic safety update report is one reporting output for devices to which the relevant requirement applies. Establish the information flow before choosing the report format.


Decision table: match the gap to the evidence


Situation

Decision

Evidence to retain

Continuing information collection

Define sources, responsibilities and review routes.

PMS plan and documented analysis.

Class I under EU MDR

Check Article 85 PMS-report obligations.

PMS report rather than an assumed PSUR.

Class IIa, IIb or III under EU MDR

Check Article 86 PSUR obligations and frequency.

PSUR appropriate to the actual device scope.


Build a usable collection and review process


Identify the sources available to your organization, such as complaints, service experience, literature and clinical follow-up. Define collection frequency, responsibilities and how records are linked to the product or family. Use consistent definitions so changes in coding or exposure do not masquerade as changes in performance.


Connect analysis with action


Review information for new risks, changes in known risks and evidence relevant to existing conclusions. Route individual reportability questions promptly through the reporting process. Record whether findings call for investigation, CAPA, updated clinical evaluation or changed surveillance activities. Where a PSUR is required, organize the relevant period’s data and analysis around justified conclusions, not a raw compilation of numbers.


Worked example: a fictional planning decision


A fictional product family has more complaints this year after expansion into a new market. The analyst preserves counts by model, issue type and exposure, rather than blending all products into one rate. A new serious case is routed immediately for reportability assessment while the periodic analysis continues. The annual reporting schedule is not used to delay case-level action.


Implementation record: make the decision reviewable


For a practical collection plan, record the source, covered device, interval, data owner, coding rules, denominator limitations and escalation route. Distinguish new information from repeated records of the same event. In the periodic output, state the period and configurations covered, relevant findings, limitations, conclusions and resulting actions. Link updates to clinical evaluation, PMCF, risk management and CAPA where relevant. Explain gaps such as incomplete distribution or exposure information rather than creating unjustified precision.


Applicability, exceptions and conclusion limits


A favorable trend does not establish that every individual event is nonreportable, and an isolated serious event may warrant action even without a rate increase. Device grouping should be justified. Regulatory frequency and submission arrangements depend on the applicable route; do not select one universal schedule for all devices or apply MDR device rules automatically to IVDs.


Choose documentation by the missing workflow


The PMS Procedure supports surveillance planning and includes a sample PMS plan. The PSUR Procedure is separately listed. The clinical bundle includes PMS but its five-product content list does not include PSUR. If your collection and reporting structures already cover the work, investigate missing sources, inconsistent coding or incomplete analysis before buying more formats.



Use the PMS procedure to organize collection, evaluation and links to risk management, clinical evaluation and quality action.



Use the PSUR procedure and reporting documents when preparing the applicable periodic safety update output.


Periodic reporting outside the EU


The PSUR is an EU MDR output. Other markets set their own periodic duties, and a surveillance plan for a device sold in several countries should list each one rather than assume the EU schedule covers them.


Market

Periodic post-market duty

Who holds it

Canada

Summary reports every 24 months for Class II devices and every 12 months for Class III and IV devices, with a concise critical analysis and a conclusion on whether the benefit-risk has changed. Health Canada is notified within 72 hours of concluding that it has. Records are kept for 7 years (CMDR sections 61.4 to 61.6).

Manufacturer

Australia

A post-marketing system that systematically reviews post-production experience, with the matters in Schedule 3 Part 1 clause 1.4 of the Therapeutic Goods (Medical Devices) Regulations reported to the TGA or the sponsor.

Manufacturer and Australian sponsor

Brazil

Post-market surveillance integrated into the quality system, with a professional designated as responsible for technovigilance (RDC 67/2009, Articles 5 and 6).

Brazilian registration holder, supplied by the manufacturer

Japan

Post-market safety management under the Marketing Authorization Holder's obligations; the manufacturer supplies the information the MAH needs.

Marketing Authorization Holder

United States

No routine periodic safety report for most devices. Post-market duties run through complaint handling, medical device reporting and, when FDA orders it, section 522 post-market surveillance studies.

Manufacturer


Canada's summary report is the closest equivalent to a PSUR, and the same analysis can usually feed both if the data are kept by market. If you sell in the MDSAP countries, the MDSAP Upgrade Kit (ISO 13485 edition or Dual-Standard edition) revises the Post Market Surveillance Procedure for these duties, and the MDSAP-ready Manufacturing Plus systems (ISO 13485 or Dual-Standard) include the revised procedure.


Common mistakes and an additional practical check


Avoid assuming every device has the same PSUR obligation or update interval. Determine the applicable requirements from device classification and the current regulatory framework, and document the rationale.


A higher complaint count after sales double may mean something different from the same increase with stable sales. Examine event severity, timing and relevant exposure before interpreting the trend; an important individual event still needs prompt assessment.


Review checklist before approving the workflow


  • Identify products, markets and applicable reporting outputs.

  • Define data sources and collection owners.

  • Document coding and exposure measures.

  • Assess findings and escalation needs.

  • Record conclusions and resulting actions.


Inside the supplied document: the controlled PMS plan


Source: supplied A.P. Lyon SOP 8-012, revision 2, effective September 19, 2026. This is a selected preview. Image wording is retained from the source; the explanation below is adapted. It is not an executed record or a quotation from a regulation.


The PMS procedure treats surveillance as a defined collection-and-review process. The selected plan instruction exposes the operating decisions behind a useful plan.


Original supplied SOP 8-012 excerpt on the controlled PMS plan

Selected original document text: the controlled PMS plan.



  • Specify sources, collection methods, owners, review frequency and analysis methods.

  • Define measurable indicators, thresholds, observation periods and comparison groups appropriate to the device.

  • Identify outputs, approvals and escalation triggers, then tailor the accompanying fields to the device or justified group.


Buyer decision: Choose the PMS procedure when the missing work is the ongoing surveillance process. A PSUR organizes a periodic evaluation for its applicable scope; PMCF addresses clinical follow-up. These documents depend on collected information and do not become evidence merely because their sections are filled.


Scope and how to use this guide


This is general implementation guidance for selecting and organizing documentation. The worked planning scenarios are fictional. Separately labeled document extracts come from the supplied A.P. Lyon files identified in their source notes. Neither is a completed study or evidence of compliance. Apply the requirements for your device, process and markets, and have the responsible technical and quality personnel approve decisions. Product downloads supply editable documentation; your organization remains responsible for its implementation and evidence.


Related buyer decisions





Primary references and review date


Technical references and linked product scope checked October 13, 2026. Regulatory summaries are identified above; checklists, decision tables and fictional examples are A.P. Lyon implementation recommendations. Consult the full applicable standards and regulations before approving a project.






Publisher and editorial standards


Published by A.P. Lyon. Aaron Lyon is the company’s founder; his professional background is described on the About page. These guides use AI-assisted research and drafting. Each guide is reviewed and approved by Aaron Lyon before publication; that review does not certify applicability to your device.




Frequently asked questions


Do all devices need a PSUR under the EU MDR?


No. Under the EU MDR, Class I devices need a post-market surveillance report (Article 85), while Class IIa, IIb and III devices need a periodic safety update report (Article 86). Confirm the device's classification before choosing the report structure.


What information should feed post-market surveillance?


Typical sources include complaints, service experience, literature and clinical follow-up. Define collection frequency, responsibilities and coding rules so that changes in coding or exposure are not mistaken for changes in device performance.


Can serious cases wait for the periodic report?


No. Individual cases that may be reportable must be routed promptly through the reporting process. The periodic PMS or PSUR schedule must never delay case-level assessment or action.


Is Canada's summary report the same as an EU PSUR?


No, but they are close. Both are periodic analyses with a benefit-risk conclusion. Canada requires one every 24 months for Class II devices and every 12 months for Class III and IV, and a notice to Health Canada within 72 hours if the benefit-risk has changed. Keeping post-market data by market lets one analysis support both.


Building a complete quality system? Build Your QMS lets you choose ISO 13485 + FDA QMSR, Dual-Standard or ISO 9001:2026 procedures in one place, and applies volume discounts automatically: 20% off 5–9 procedures, 35% off 10–19, 50% off 20–39 and 60% off 40 or more.


Not sure where your quality system stands? Take the free QMS gap check: about 20 questions, an instant report of your gaps by clause, and the procedures that close each one.


Just getting started? The free Startup Quality Plan shows which procedures you need now, which can wait, and how long they take to put in place.