ISO 11607 Packaging Validation: What the Standard and the 2023 Amendments Require
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 24, 2026.
Many device companies think of sterile packaging validation as sealer OQ and PQ plus an aging study. Those are important, but they are only part of what ISO 11607 asks for. An auditor who opens your packaging file will also look for a packaging risk management process, qualified materials with microbial barrier evidence, a usability evaluation for aseptic presentation, distribution performance testing, validated test methods and a recorded decision every time something changes.
This guide walks through what ISO 11607-1:2019 and ISO 11607-2:2019, each with Amendment 1:2023, require for a terminally sterilized device, in the order the work is usually done. Each summary is in our own words.
Why ISO 11607 applies to you
ISO 13485:2016 clauses 7.5.5 and 7.5.7 require processes for sterilization and sterile barrier systems to be validated under documented procedures before they are used, and after product or process changes as appropriate. Since FDA's Quality Management System Regulation took effect, a US device manufacturer must document a quality management system that complies with the applicable requirements of ISO 13485 (21 CFR 820.10(a)). ISO 11607 is the standard that sets out how to validate the sterile barrier system and the packaging system around it.
The standard comes in two parts:
ISO 11607-1 covers materials, sterile barrier systems and packaging systems: what they are made of, how they are designed, and how you show they keep the device sterile until it is opened.
ISO 11607-2 covers the forming, sealing and assembly processes that make the sterile barrier system: process specification, installation, operational and performance qualification, routine control and change.
Two terms to get right first
ISO 11607-1 Annex E separates two things that are often blurred together. The sterile barrier system is the minimum package that keeps microorganisms out and allows the product to be presented aseptically, for example a pouch or a sealed tray. Protective packaging is everything around it that protects the sterile barrier system during handling, storage and distribution, for example a shelf carton and a shipper. The two together make the packaging system.
The distinction matters because the requirements differ. The sterile barrier system must keep its integrity until the point of use. The protective packaging must protect that integrity. When you test, you test the sterile barrier system for integrity after putting the whole packaging system through distribution and aging.
Packaging risk management: the biggest change in 2023
Amendment 1:2023 to ISO 11607-1 changed clause 4.2. It now requires a packaging risk management process that meets a new normative Annex F, and clause 6.1.1 now requires the packaging system to be designed to minimize the Annex F risks to the user and the patient during intended use and reasonably foreseeable misuse.
Annex F describes an ongoing, documented process that covers design and development, validation, production and post-production. In practice that means:
a risk management plan for each packaging system, with documented risk acceptability criteria;
hazard identification under normal and fault conditions;
risk estimation and evaluation against the criteria;
risk control, with verification of each control measure;
review of production and post-production information, with corrective and preventive actions.
Amendment 1:2023 to ISO 11607-2 added the same idea for the processes, in clause 4.2 and a new normative Annex B. Each forming, sealing and assembly process has a risk management plan, which may be combined with the device risk plan or shared across similar processes when the rationale is documented. The hazards to analyze are microbial contamination, chemical contamination, adverse environmental and processing conditions, and misleading information. Risk control follows a set order: first process development or a design change, then process controls such as monitors, in-process checks, alarms and fixtures, and only then information to operators and inspection of the output.
If your packaging file has no risk plan, this is the first gap to close.
Materials and preformed sterile barrier systems
ISO 11607-1 clause 5 sets the requirements for the materials and for preformed sterile barrier systems such as pouches, header bags and lidded trays you buy ready to seal. Materials are selected to meet the goals of the packaging system, and the conditions they are made and handled under are set and controlled where applicable. Their source, history and traceability must be known, especially for recycled material. Conformity to an EN 868 part on its own is not full conformity with ISO 11607-1.
Clause 5.1.6 lists the properties to evaluate: microbial barrier; biocompatibility and toxicology, usually for material that touches the device; physical and chemical properties; compatibility with the forming, sealing and assembly processes; compatibility with the sterilization process; and any use-by limit before sterilization and shelf-life limit after it.
Two of these need particular evidence:
Microbial barrier (clause 5.2). A material not declared porous is shown impermeable with the Annex C air-passage test. A porous material, such as medical paper or spunbonded polyolefin, must give an adequate microbial barrier. A validated physical test that correlates with a validated microbiological challenge method is acceptable evidence. ASTM F1608-21 is one exposure-chamber method that ranks porous materials by log reduction value. It compares materials; it does not predict how a material performs in your specific package.
Sterilization compatibility (clause 5.3). Materials and sterile barrier systems are shown suitable for the specified sterilization process, cycle parameters and limits, including repeated or different exposures where they are specified.
Labelling is part of this too. Under clause 5.4, labels stay attached, intact and legible until the point of use, survive sterilization without affecting it, and use no ink that can transfer to the device or fade until illegible.
Designing the packaging system
ISO 11607-1 clause 6.2 requires documented procedures for packaging system design and development. The design considers the user and the use environment, the mass and shape of the product, sharp edges or protrusions, protection needs, the product's sensitivity to radiation, moisture, shock or static, the number of items per package, labelling, environmental limits, expiry, distribution, handling and storage, and sterilization compatibility and residuals.
When similar devices share a packaging system, clause 6.1.9 lets you validate a family. The rationale for the similarity and for picking the worst-case configuration must be recorded, and at least the worst case must be tested. Clause 8.2.3 adds that the worst case for performance testing takes account of all specified sterilization exposures and the most challenging contents.
Usability evaluation for aseptic presentation
Clause 7 requires a documented usability evaluation. It answers three questions: can the user tell where to start opening, can the user recognize and perform the opening technique without contaminating or damaging the contents, and can the user then present the contents aseptically? It can be done under real or simulated use, and it can be shared across product and packaging families on a worst-case or other valid rationale. If any of the three is not met, the sterile barrier system is redesigned or the instructions are improved, and the evaluation is repeated.
This is a requirement many files are missing, because it is not a laboratory test. It needs a protocol, representative users and a record.
Performance and stability testing
Clause 8 separates two kinds of testing, and they must stay separate:
Performance testing shows the packaging system protects the sterile barrier system through handling, distribution and storage hazards such as shock, vibration, compression, temperature, humidity and pressure changes. It uses the worst-case sterile barrier system in the worst-case protective packaging. The protective packaging is adequate when the sterile barrier system passes integrity testing afterwards.
Stability testing shows the sterile barrier system keeps its integrity over time and uses real-time aging. Accelerated aging is acceptable evidence for an expiry claim until real-time data exist. Real-time aging starts before commercial release, the accelerated study starts within three months of the real-time study unless another rationale is documented, and the accelerated conditions and duration need a written rationale. ASTM F1980-21 gives the method most companies use to calculate accelerated aging time.
In both cases integrity is tested on sterilized samples after the conditioning. A useful point from clause 8.2: new sealing equipment with equivalent technology does not need new performance testing if its process is validated and makes seals that meet the original package specification.
Test methods, sampling and records
Clause 4 applies to every test in the file:
Sampling (4.3). Sampling plans fit the item being tested and rest on a statistically valid rationale.
Test rationale (4.4.1 and 4.4.2). The reason for choosing each test, and the reason for each acceptance criterion, are written down. Pass or fail counts as an acceptance criterion.
Test method validation (4.4.3 and 4.4.4). Every test method used to show conformity is validated and documented by the laboratory that runs it, including repeatability, reproducibility and, for integrity tests, sensitivity. The 2023 note is plain: a method published by a standards body is not validated for you just because it is published. Annex B lists methods to choose from or justify against.
Records (4.5). Conformity is recorded, records are kept for a set period that reflects the device's expiry and traceability, and electronic records and signatures used for validation stay legible and retrievable.
Process validation under ISO 11607-2
The sealing, forming and assembly processes are validated with IQ, OQ and PQ, in that order, against a process specification that states the required output, the variables monitored and the parameters controlled. The final step is a documented review and formal approval of the validation. Routine monitoring then keeps the process within the validated parameters. We cover this in detail in Sealer IQ, OQ and PQ Under ISO 11607-2.
When the packaging system is validated, and when to revalidate
Under ISO 11607-1 clause 9, a packaging system is validated when the design, usability, performance and stability requirements are met and its sterile barrier system conforms to ISO 11607-2. Packaging design documents are under change control.
Revalidation is needed when a change to the design, contents, materials or configuration compromises the validation and can affect sterile barrier integrity, or when storage, handling or distribution hazards exceed what was tested. Examples include new materials, raw material changes, sterilization changes, new shipping configurations and market feedback about integrity. The need for revalidation is evaluated and recorded, and the revalidation may be narrower than the original. Minor changes are recorded too, because several minor changes can add up.
Finally, clause 10 asks that, where user instructions are appropriate, they tell the user to inspect the packaging for integrity before use.
A typical order of work
Write a packaging validation master plan that names the packaging system, the family and worst case, the tests and the acceptance criteria.
Prepare the packaging risk management plan and the first risk assessment.
Qualify the materials and preformed sterile barrier systems, including microbial barrier and sterilization compatibility evidence.
Validate the test methods you will use.
Validate the sealing and forming processes with IQ, OQ and PQ.
Run the usability evaluation.
Build sterilized samples on the validated process, then run performance testing and start the accelerated and real-time aging studies.
Write the validation summary report and approve it.
Put routine lot testing and change assessment in place.
How the A.P. Lyon kit fits
The Sterile Packaging Validation Kit has 5 procedures and 11 forms for every step above: the master plan, the packaging risk management plan and assessment, material qualification, the sealer installation qualification protocol, test method validation, the usability evaluation, performance testing, routine lot testing, change assessment and the summary report, plus a requirements matrix that maps each ISO 11607 requirement to the document that meets it. For the sealer OQ and PQ test work, use our bar sealer, tray sealer or 4-side seal pouch protocols. For aging, use the Accelerated Aging Study Protocol.
Related reading: Seal Strength, Dye Penetration or Visual Inspection: Choosing Package Tests and Medical Device Shelf Life and Accelerated Aging: Define the Claim First.
Frequently asked questions
What changed in ISO 11607 Amendment 1:2023?
The main change is risk management. ISO 11607-1 clause 4.2 now requires a packaging risk management process that meets the new normative Annex F, and ISO 11607-2 adds a normative Annex B for forming, sealing and assembly processes. The amendments also make clear that a published test method is not validated for the user just because it is published.
Is sealer OQ and PQ enough to validate sterile packaging?
No. Process validation covers the sealing process under ISO 11607-2. ISO 11607-1 also requires material qualification, a design procedure, a usability evaluation, performance and stability testing, validated test methods and change control for the packaging system.
Can accelerated aging support an expiry date on its own?
Only until real-time data exist. ISO 11607-1 clause 8.3 accepts accelerated aging as evidence for an expiry claim, but real-time aging must also be done and must start before commercial release.