Supplier Quality Agreement for Medical Device Components: What to Put In It
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 18, 2026.
Sooner or later, every supplier to the medical device industry is handed a quality agreement to sign. Some arrive as a 30-page customer document full of clauses written for a different kind of supplier. Others arrive as a one-page form that leaves the hard questions blank. Either way, the agreement will govern how you handle changes, nonconformances, records and audits for years, so it pays to understand what belongs in it and what you are promising.
This guide is written from the supplier's side. It explains why device customers need a quality agreement, the sections a good one contains, and the points to settle before you sign.
Why device customers need a written agreement
A device manufacturer that is subject to FDA's Quality Management System Regulation must document a quality management system that complies with ISO 13485 (21 CFR 820.10(a)). ISO 13485:2016 clause 4.1.5 says that when the manufacturer outsources a process that affects product conformity, it keeps responsibility for that process, controls it in proportion to the risk and to the supplier's ability, and includes written quality agreements in those controls.
Clause 7.4.2 adds a specific item: purchasing information includes, as applicable, a written agreement that the supplier notifies the manufacturer of changes before implementing changes that affect the product's ability to meet purchase requirements.
So for many suppliers the quality agreement is not optional paperwork. It is how the customer shows its auditor that it controls you. The component supplier itself is not covered by Part 820 (21 CFR 820.1(a)(2)), so the agreement is the main place where the device rules reach your shop.
Quality agreement versus purchase order
The purchase order and commercial terms cover price, quantity, delivery, payment and liability. The quality agreement covers who is responsible for each quality activity and how the two companies communicate about quality. Keep them separate. A good agreement says that it supplements, and does not replace, the commercial terms, and that where the agreement and a purchase order differ on a quality matter, the agreement governs.
The agreement also lasts longer than any single order. It should have a defined review interval, such as every two years or after a major change, and an agreement number and revision so both sides know which version is current.
The sections a good agreement contains
1. Parties and contacts
Legal names, sites, a quality contact on each side and, importantly, a named contact for change notices. Many change notices go astray because they are sent to a buyer who has moved on. List the certifications you hold, with certificate numbers and expiry dates.
2. Scope
The part numbers, revisions and descriptions covered, the processes you perform, and the site where you perform them. Two questions belong here because they drive the rest of the agreement: does the customer state that the part is used in an implantable device, and does the device support or sustain life? For those devices, ISO 13485 clause 7.5.9.2 traceability applies, and FDA applies the same clause to life-supporting and life-sustaining devices (21 CFR 820.10(d)). Record the traceability level and any cleanliness specification number in the scope too.
3. Responsibilities
This is the heart of the agreement: a table of activities with C (customer), S (supplier) or C/S (shared) against each, and a column for agreed times and values. A useful table covers at least:
drawings and specifications, and working only to the current revision
classification of the finished device and all regulatory submissions (always the customer)
maintaining the supplier's quality management system
review of flow-down requirements before accepting each order
advance written notice of changes, with the lead time
which changes need customer approval before changed product ships
control of sub-tier suppliers and flow-down of these requirements
process validation where output is not fully verified, and who approves it
first article inspection for new parts and changes
cleanliness specification and test methods, and the cleaning process and its validation
traceability records to the agreed level
label content, including any unique device identifier (the customer's responsibility)
inspection, release and a certificate of conformance with each shipment
no shipment of known nonconforming product without a written concession
notice of nonconformance found after shipment, with a time limit
response times for supplier corrective action requests
support for complaint and failure investigations
decisions on field actions, recalls and regulatory reports (always the customer), and the time you have to supply records for them
protection of customer property such as molds, fixtures and material
record retention, and no destruction of records without notice
customer audits of your site and records
notice to the customer if a regulatory authority contacts you about the customer's device
preservation, packaging, storage and shipping
confidentiality of drawings and data
4. Term, review and signatures
Effective date, review interval, next review date, and signatures from quality and purchasing on the customer side and quality and management on the supplier side.
Points to settle before you sign
Change notice lead time and scope
The agreement should say how far in advance you must notify the customer and which changes need its approval. Read the list carefully. If it says "any change," ask for it to be narrowed to changes that could affect the part's ability to meet requirements, which is the test in ISO 13485 clause 7.4.2, plus any specific items the customer cares about, such as material source, sub-tier special processes, manufacturing site and validated process parameters. Make sure the lead time is one you can meet for changes at your own suppliers, because you are passing the same obligation down to them.
Record retention
ISO 13485 clause 4.2.5 requires the manufacturer to keep records at least for the lifetime of the device as it defines it, or as regulations require, and not less than two years from release. The customer will ask you for a matching period. Get a number in years written into the agreement, check it against your own retention schedule, and agree what happens at the end: many agreements require you to offer the records to the customer before destroying them.
Response times
Notice of a nonconformance found after shipment, containment on a corrective action request, a full corrective action response, investigation support and supply of records for a field action all need times. Agree times you can actually meet on a bad day, not on a good one. A missed deadline in the agreement is itself a finding in the customer's next supplier audit.
Validation and first article
If you perform a process whose output cannot be fully verified by inspection, such as welding, molding or cleaning, ISO 13485 clause 7.5.6 requires it to be validated. Settle who writes the protocol, who approves it and who pays for revalidation after a change. The same applies to first article inspection after a change.
Cleanliness
If the drawing says only "clean and free of contaminants," ask for the cleanliness specification and test methods. Under ISO 13485 clause 7.5.2, cleanliness requirements must be documented in specific situations, including when process agents must be removed from the product. The customer should define what clean means; you should define and validate how you achieve it.
Regulatory contact and confidentiality
If FDA or another regulator ever contacts you about the customer's device, the customer will want to know at once. A short limit, such as one business day, is a reasonable commitment. In return, confidentiality of drawings and process data should be mutual.
Common mistakes
Signing the customer's document without reading the responsibilities. Long agreements often assign the supplier activities it does not do, such as sterilization or labeling.
Leaving the time and value columns blank. A blank lead time or retention period is an argument waiting to happen.
No change notice contact. Notices go to a buyer's inbox and nobody at the customer reviews them.
No flow-down to sub-tier suppliers. You promise advance notice of changes, but your plating house changes chemistry without telling you.
No internal procedures behind the promises. The agreement says you will notify, trace and hold, but your quality system has no procedure that makes it happen. The customer's auditor will ask to see the procedure.
Making the agreement work in your shop
An agreement only helps if your quality system carries it out. Each commitment should point to a procedure: contract review to capture the agreement's requirements on each order, change notification, traceability, cleanliness, and concession and nonconformance notification. Train the people who quote, buy, change and release product on what the agreement says.
The Supplying the Medical Device Industry supplier pack includes a Supplier Quality Agreement for Medical Device Customers (FORM 0101-MDS) with the responsibilities table described above and default assignments as a starting point for discussion, plus the five procedures that carry out each commitment. It adds to the ISO 9001 QMS Complete system. Device manufacturers writing agreements for their own suppliers can also read Medical Device Quality Agreements and Consulting Contracts.
Related reading: ISO 13485 Requirements for Suppliers: What Device Customers Pass Down and Change Notification: What Medical Device Customers Expect from Suppliers.
Frequently asked questions
Is a supplier quality agreement required by ISO 13485?
For outsourced processes, yes: clause 4.1.5 says the manufacturer's controls include written quality agreements. For purchased product, clause 7.4.2 calls for a written agreement on change notification, as applicable. Customers usually cover both in one quality agreement.
Who decides on a recall when a supplier's part is defective?
The device manufacturer. A good agreement gives decisions on field actions, recalls and regulatory reports to the customer, and gives the supplier a time limit to provide the records and data needed to support them.
Should the quality agreement replace the purchase order terms?
No. The agreement covers quality responsibilities and supplements the commercial terms. It should say that it governs quality matters where it and a purchase order differ.