What Documents Does ISO 13485 Require? The Complete List
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 16, 2026.
ISO 13485 is specific about documentation. It asks for a quality manual, a file for every device type or family, a long list of documented procedures and the records that prove the system is running. Clause 4.1.1 then adds a catch-all: anything the standard or the regulations you operate under require to be documented must be established, implemented and maintained.
That makes "what do we actually need?" a fair question, especially for a first certification. This guide answers it clause by clause, using ISO 13485:2016, and then covers the few extra documentation requirements in the FDA's Quality Management System Regulation (QMSR, 21 CFR Part 820) for companies that sell in the United States.
Three kinds of documentation
It helps to sort what ISO 13485 asks for into three groups:
Top-level documents. The quality policy, quality objectives, quality manual and the medical device file. These describe the system and the product.
Documented procedures and arrangements. The written way you carry out a process, such as complaint handling or internal audit. Some clauses say "documented procedure"; others say requirements or arrangements must be "documented". Both mean you need it in writing and under document control.
Records. The evidence that you did what the procedures say. ISO 13485 names many of them, and clause 4.1.3 asks for records that show conformance to the standard and compliance with regulations for every QMS process.
Top-level documents
Quality policy and quality objectives (4.2.1, 5.3, 5.4.1). The policy commits to complying with requirements and keeping the QMS effective. Objectives must be measurable, consistent with the policy and include those needed to meet regulatory and product requirements.
Quality manual (4.2.2). It states the scope of the QMS, gives details of and the justification for any exclusion or non-application, includes or references the documented procedures, describes how the processes interact and outlines the documentation structure.
Your regulatory role (4.1.1). Document the role or roles the organization takes on, for example manufacturer, authorized representative, importer or distributor. The quality manual is a natural place for it.
Responsibilities and authorities (5.5.1). Defined, documented and communicated, including how the people who manage, perform and verify work affecting quality relate to each other. An organization chart plus a responsibility section in the manual is a common way to cover it.
Medical device file (4.2.3). One or more files for each device type or family. Each contains or references the general description, intended use, labelling and instructions for use, product specifications, the specifications or procedures for manufacturing, packaging, storage, handling and distribution, measuring and monitoring procedures, and, as appropriate, installation requirements and servicing procedures.
Design and development file (7.3.10). Where you design devices, a design and development file is kept for each device type or family.
Documented procedures and arrangements, clause by clause
Clause 4: Quality management system
Software used in the QMS (4.1.6). Documented procedures for validating software used in the QMS, before first use and after changes as appropriate, with an approach proportionate to the risk of its use.
Outsourced processes (4.1.5). Controls proportionate to the risk, including written quality agreements with the external party.
Document control (4.2.4). Changes are reviewed and approved by the original approving function, or another designated function with access to the background information. Keep at least one copy of obsolete documents for a defined period that covers at least the device lifetime and is not shorter than the retention period of any record made from them.
Control of records (4.2.5). A documented procedure covering identification, storage, security and integrity, retrieval, retention time and disposition of records, including protection of confidential health information where regulations require it.
Clause 5: Management responsibility
Management review (5.6.1). A documented procedure, with reviews at documented planned intervals.
Clause 6: Resource management
Competence, training and awareness (6.2). The process for establishing competence, providing training and ensuring awareness is documented.
Infrastructure (6.3). Documented infrastructure requirements, and documented maintenance requirements, including intervals, where maintenance can affect product quality.
Work environment (6.4). Documented work environment requirements and the procedures to monitor them where conditions can affect quality. For sterile devices, documented requirements for control of microbial and particulate contamination.
Personnel health, cleanliness and clothing (6.4.1). Documented where contact with the product or work environment could affect device safety or performance.
Contaminated product (6.4.2). Arrangements to control contaminated or potentially contaminated product, planned and documented as appropriate.
Clause 7: Product realization
Risk management (7.1). One or more documented risk management processes for product realization, with records kept. The output of realization planning is documented as well.
Communication with customers (7.2.3). Documented arrangements for product information, enquiries, contracts and orders, customer feedback including complaints, and advisory notices.
Design and development (7.3.1, 7.3.8). Documented design and development procedures, with design transfer carried out by documented procedure.
Production (7.5.1). Documented production procedures as appropriate, as part of planned and controlled production and service provision.
Product cleanliness (7.5.2). Documented cleanliness or contamination control requirements in the cases the clause lists, for example when you clean product before sterilization or use.
Installation (7.5.3). Documented installation requirements and acceptance criteria for verifying installation, as appropriate.
Servicing (7.5.4). Where servicing is a specified requirement, documented servicing procedures, reference materials and reference measurements as needed.
Process validation (7.5.6). Documented procedures for validating processes whose output is not or cannot be verified, covering approval criteria, equipment and personnel qualification, methods, acceptance criteria, statistical techniques with a rationale for sample size, revalidation criteria and approval of changes.
Sterilization and sterile barrier systems (7.5.7). Validation under documented procedures, before implementation and after product or process changes as appropriate.
Unique device identification (7.5.8). Where regulations require it, a documented system for assigning UDI.
Traceability (7.5.9.1). Documented traceability procedures that set the extent of traceability and the records to keep.
Preservation (7.5.11). Documented procedures for preserving product conformity during processing, storage, handling and distribution.
Clause 8: Measurement, analysis and improvement
Feedback (8.2.1). A documented feedback process that gathers data from production and post-production activities.
Complaint handling (8.2.2). Documented procedures for timely complaint handling, from receipt through investigation, the decision on reporting to regulatory authorities and the need for correction or corrective action.
Reporting to regulatory authorities (8.2.3). Where regulations require notification, documented reporting procedures.
Internal audit (8.2.4). A documented procedure, with the audit criteria, scope, interval and methods defined and recorded. Auditors do not audit their own work.
Product release (8.2.6). Monitoring and measurement of product at the applicable stages under planned, documented arrangements.
Advisory notices (8.3.3). Documented procedures for issuing advisory notices that can be put into effect at any time.
Rework (8.3.4). Rework follows documented procedures that consider possible adverse effects.
Analysis of data (8.4). Documented procedures to determine, collect and analyse data, with input at a minimum from feedback, product conformity, process and product trends, suppliers, audits and, as appropriate, service reports.
Corrective action (8.5.2). A procedure covering review of nonconformities including complaints, cause determination, action planning, verification that the action does not adversely affect regulatory compliance or device safety and performance, and review of effectiveness.
Preventive action (8.5.3). A documented procedure for potential nonconformities, from cause through effectiveness review.
Records to keep as evidence
ISO 13485 names these records directly:
Management review (5.6.1, 5.6.3). Records of each review, with outputs recording decisions on improvement, product, regulatory changes and resource needs.
Training and competence (6.2). Records of competence and training, and of the evaluation of whether the training was effective.
Maintenance (6.3). Records of maintenance where maintenance can affect product quality.
Risk management (7.1). Records of the risk management activities for product realization.
Review of product requirements (7.2.2). Records of the review carried out before committing to supply.
Supplier evaluation and monitoring (7.4.1). Records of supplier evaluation, selection and monitoring results.
Device or batch record (7.5.1). For each device or batch, a record that gives traceability, identifies the amount manufactured and the amount approved for distribution, and is verified and approved.
Installation and servicing (7.5.3, 7.5.4). Records of installation and its verification, and of servicing, whether done by you or your supplier. Servicing records are analysed to decide whether the information is a complaint.
Sterilization (7.5.5). Process parameters for each sterilization batch, traceable to each production batch.
Process validation (7.5.6, 7.5.7). Validation records, as the documented validation procedures require.
Traceability for implantable devices (7.5.9.2). Components, materials and work environment conditions that could affect safety and performance, and the name and address of the shipping package consignee.
Customer property (7.5.10). Records of any loss, damage or unsuitability reported to the customer.
Calibration (7.6). Calibration and verification results, and the assessment of the validity of earlier results when equipment is found nonconforming.
Complaints (8.2.2). Complaint records, including a documented justification when a complaint is not investigated.
Internal audits (8.2.4). Audit records, including the criteria, scope, interval and methods.
Product release (8.2.6). Evidence of conformity to acceptance criteria and the identity of the person who authorized release. For implantable devices, the identity of the people who performed any inspection or testing.
Concessions (8.3.2). The justification and approval for accepting nonconforming product, with the person who authorized it identified.
Data analysis (8.4). Recorded results.
How long to keep them
Clause 4.2.5 sets the floor: keep records for at least the device lifetime defined by your organization, or longer if a regulation requires it, and never less than two years from the date of product release. Changes to records must remain identifiable.
What the FDA QMSR adds
If you sell devices in the United States, the QMSR (21 CFR Part 820) works from ISO 13485 and adds a short list of documentation requirements of its own:
UDI in complaint and service records (820.35(a) and (b)). Complaint records and service records must include the unique device identifier (UDI) or universal product code (UPC) where one applies.
UDI in the production record (820.35(c)). The UDI is recorded for each device or batch of devices.
Labeling inspection and release (820.45). Labeling is inspected for accuracy before release or storage, including the UDI or other identification, expiration date and storage and handling instructions. The release and the inspection results are documented under your records procedure, and labeling and packaging operations are controlled to prevent mix-ups.
Complaints that are reportable (820.10(b)(3)). Your complaint handling must notify FDA of complaints that meet the Part 803 reporting criteria, and the records for those complaints must include the information listed in 820.35(a).
Records FDA can read (former 820.180(c) removed). Management review, internal audit and supplier audit records are no longer exempt from FDA inspection. Procedures must not say these reports are withheld from FDA, and the records should be ready for an investigator to read.
If your management review procedure or quality manual still says those reports are not shown to FDA, it needs updating.
Putting the set together
A few practical points:
Start from the manual. Writing the scope, any non-applications and the process map first tells you which of the procedures above apply. A company that does no servicing or installation, or makes no sterile devices, justifies those as non-applications in the manual instead of writing procedures for them.
Keep a master list with a clause cross-reference. Auditors often ask where a clause is covered. A list that maps each document to its clauses answers that in seconds and shows you any gaps.
Write procedures people will follow. ISO 13485 cares that the system is documented and effective (4.1.1). A short procedure that matches how the work is really done holds up better than a long one nobody reads.
Plan the records up front. An auditor checks the evidence, not only the procedure. Build the forms with the procedures so the records are created as the work is done.
Get the documents in place
A.P. Lyon's ISO 13485 QMS packages are written to ISO 13485:2016 and the FDA QMSR, as editable Microsoft Word and Excel files:
ISO 13485 QMS Base: quality manual plus 23 procedures with related forms, $1,195.
ISO 13485 QMS Design Plus: quality manual plus 45 procedures with related forms, $1,995.
ISO 13485 QMS Manufacturing Plus: quality manual plus 73 procedures with related forms, $2,795.
Already have a system written to the old Quality System Regulation? The QMSR Gap Kit ($495) walks you through the update.
Need only some of the documents? Choose individual procedures in Build Your QMS. Discounts apply automatically: 20% off at five procedures, rising to 60% off at forty or more.
Not sure what you are missing? Take the free QMS gap check: about 20 questions, an instant report of your gaps by clause, and the procedures that close them.
Just getting started? The free Startup Quality Plan shows which procedures you need now, which can wait, and how long they take to put in place.
Need a budget for management? The free QMS Budget Planner compares the cost and time of writing your system in-house, hiring a consultant, adapting A.P. Lyon documents or ordering a Custom QMS, and gives you a one-page memo for your CEO or CFO.
Related guides: ISO 9001 vs ISO 13485: Which Do You Need? and What Documents Does ISO 9001:2026 Require?