Seal Strength, Dye Penetration or Visual Inspection: Choosing Package Tests
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 24, 2026.
Three package tests show up in almost every sterile packaging file: seal strength to ASTM F88, dye penetration to ASTM F1929 and visual inspection to ASTM F1886. They are often treated as interchangeable checks, or all three are run on every sample "to be safe." They are not interchangeable. Each one measures something different, has limits on what it can detect, and must be validated for your package before its results count.
This guide explains what each test tells you, where it fits in ISO 11607 validation and routine production, and what the test record needs. Each summary is in our own words from ASTM F88/F88M-23, ASTM F1929-23, ASTM F1886/F1886M-25 and ISO 11607-1 and -2:2019 with Amendment 1:2023.
What ISO 11607 asks of any package test
Before choosing a test, it helps to know the rules that apply to all of them:
Rationale. The reason for choosing each test, and the reason for each acceptance criterion, are written down. Pass or fail counts as an acceptance criterion (ISO 11607-1 and -2, clauses 4.4.1 and 4.4.2).
Validation. Every test method used to show conformity is validated and documented by the laboratory that runs it, covering repeatability, reproducibility and, for integrity tests, sensitivity. A method published by a standards body is not validated for you just because it is published (clauses 4.4.3 and 4.4.4, as amended in 2023).
Sampling. Sampling plans fit what is being tested and rest on a statistically valid rationale (clause 4.3).
Choice of method. Methods are chosen from, or justified against, Annex B of ISO 11607-1.
ASTM F88 and ASTM F1886 both point to ASTM F3263 for how to validate a package test method for your specific package.
Seal strength (ASTM F88): how strong is the seal?
Seal strength is a quantitative test. A strip cut across the seal is pulled apart in a tensile tester and the force per unit width is recorded. ASTM F88 describes it as a measure for process validation, process capability and process control. A minimum seal strength is a package requirement, and an upper limit may also be set so that a user can open a peelable package.
What seal strength does not do is show integrity. A seal can be strong and still have a channel running through it. That is why seal strength is always paired with an integrity test.
Key points for the procedure:
Tail-holding technique. Name one technique and use it for every test in a series: unsupported (A), supported at 90 degrees by hand (B) or supported at 180 degrees (C). The technique changes the result, so it and any fixtures are recorded.
Equipment. A constant-rate-of-jaw-separation tensile tester, with load and grip-separation measurements each accurate to within plus or minus 2 percent and calibrated as the manufacturer recommends.
Specimens. Cut with clean edges perpendicular to the seal, normally 25 mm (1 in.) wide; any other width is justified. Record where on the package each specimen came from. Do not remove specimens with visual defects without a reason.
Test speed. Grips separate at 200 to 300 mm/min (8 to 12 in./min).
Results. For each specimen record the maximum force and every failure mode: adhesive peel, cohesive peel, delamination, material break or tear, or elongation. Report average peel force only when the seal actually peels, and say how it was calculated.
Seal strength is the workhorse of OQ and PQ, because it shows where the sealing window is and how capable the process is.
Dye penetration (ASTM F1929): is there a leak path through the seal?
Dye penetration is a pass or fail integrity test for seal leaks. A dye solution is applied to the edge of the seal from inside the package, and the seal is viewed through the transparent side. Dye that passes through a defined channel to the other side of the seal is a leak. Dye that wicks into the porous material is not.
Its limits matter:
It applies to seals between a transparent material and a porous material that holds the dye for at least 5 seconds without discoloring the seal area. Clear film to medical paper or spunbonded polyolefin is the typical case.
It detects and locates channels the size of a 50 micrometre (0.002 in.) wire. It does not measure leak size.
Packages must be dry. Any package exposed to liquid is dried at normal storage temperature first.
Oxidative sterilization and folding of porous material can cause false positives, and should be evaluated.
The procedure states which method is used: injection (A), edge dip (B) or eyedropper (C). Each seal is viewed for no more than about 5 seconds per side and 20 seconds in total for a 4-sided package, optionally with 5x to 20x magnification. For injection, a flexible tube on the syringe through a cut opening is recommended instead of a hypodermic needle. The dye solution itself is defined by weight: 0.5 percent nonionic wetting agent, 0.05 percent toluidine blue and 99.45 percent water. A different dye needs its own precision and bias data.
Dye penetration is the usual integrity test after OQ, PQ, distribution testing and aging, wherever the materials allow it.
Visual inspection (ASTM F1886): is anything visibly wrong?
Visual inspection is a qualitative accept or reject check in two parts. Part 1 covers seals, looking for channels, and needs at least one transparent side. Part 2 covers material surfaces, transparent or opaque. It is non-destructive, which makes it the natural choice for 100 percent or frequent inspection in production.
Its limits are about people and conditions. Detection depends on defect size, contrast, materials, lighting, magnification, the time allowed, and the inspector's training and experience. In the interlaboratory study behind the standard, a 75 micrometre channel was found only 60 to 100 percent of the time. That is why the user must define, during test method validation, what defect size the inspection can reliably detect and which defects to look for.
Conditions and procedure:
Inspect under about 540 lux (50 foot-candles) or other suitable lighting without glare, at 30 to 45 cm (12 to 18 in.), with enough time to see the whole seal and surface. A UV light box can be used for fluorescent coatings.
Inspect the entire seal for completeness and uniformity, and the entire surface for visible defects. Mark and record the number and location of channels and suspect defects for each package.
Classify other seal and surface features against your own accept and reject criteria, and define the actions to take when they appear in production.
Confirm any suspect defect with an integrity test. If the package passes, it may be accepted. If not, it is an integrity defect. ASTM F1886 points to ASTM F2097 for choosing that test.
Choosing the right test for the question
Is the seal within its strength range? Use ASTM F88 seal strength. It is quantitative and destructive. Typical use: the OQ and PQ window, routine process control and testing after aging.
Is there a channel through the seal? Use ASTM F1929 dye penetration. It is pass or fail and destructive. Typical use: integrity after OQ, PQ, distribution testing and aging.
Is anything visibly wrong with seals or surfaces? Use ASTM F1886 visual inspection. It is accept or reject and non-destructive. Typical use: 100 percent or frequent inspection in production, and the first screen in every study.
A common combination in validation is visual inspection of every sample, then seal strength on some and dye penetration on others, because both are destructive. In routine production, visual inspection is often done on every package or at high frequency, with seal strength and dye penetration at a defined sampling frequency. Whatever you choose, ISO 11607 asks you to write down why.
Validating the test methods
Validation is where many files fall short. For each method, the laboratory that runs it documents:
Repeatability: the same operator and equipment get consistent results on the same kind of sample.
Reproducibility: different operators, shifts or equipment get consistent results.
Sensitivity, for integrity tests: the smallest defect the method reliably detects in your package. For dye penetration and visual inspection this usually means making samples with known channels and checking the detection rate.
Validate with your own materials, your seal widths and your inspectors. A method validated on a different package does not carry over automatically.
What goes in the test record
Each standard lists what to record. Common elements for all three are: material identification, the test equipment and its calibration, the operator, the date, the number of specimens and the sampling method, the results, and any deviation. Some items are specific:
ASTM F88: ambient temperature and humidity, grip separation rate and initial distance, seal width, tail-holding technique and fixtures, which material was in each grip, failure modes, maximum force (and average force with its method, if reported) and the statistics used.
ASTM F1929: the method used, conditioning, any non-standard dye, whether inspection was aided, and a description or sketch of leak sites.
ASTM F1886: lighting and inspection conditions, and the number and location of channels and suspect defects for each package, with the integrity test result for any suspect defect.
How the A.P. Lyon kit fits
The Sterile Packaging Validation Kit includes the Package Test Method Validation Procedure and its record, which cover repeatability, reproducibility and sensitivity for seal strength, dye penetration and visual inspection, and the Routine Seal Strength and Package Integrity Testing Procedure with a lot test record for production. The kit's requirements matrix maps each of these ASTM and ISO 11607 requirements to the document that meets it.
Related reading: ISO 11607 Packaging Validation: What the Standard and the 2023 Amendments Require and Sealer IQ, OQ and PQ Under ISO 11607-2.
Frequently asked questions
Does a passing seal strength test prove package integrity?
No. ASTM F88 states that seal strength is not by itself a measure of package integrity. A seal can meet its strength specification and still have a channel, so seal strength is paired with an integrity test such as dye penetration.
Can I use dye penetration on a foil or film-to-film package?
ASTM F1929 applies to seals between a transparent material and a porous material that holds the dye for at least 5 seconds without discoloring the seal area. For other material combinations a different integrity test is needed, chosen and justified against ISO 11607-1 Annex B.
Is ASTM F1886 visual inspection already validated because it is a standard?
No. ISO 11607 says a published method is not validated for you because it is published, and ASTM F1886 asks the user to define the detectable defect size and defect types during test method validation.
Building a complete quality system? Build Your QMS lets you choose ISO 13485 + FDA QMSR, Dual-Standard or ISO 9001:2026 procedures in one place, and applies volume discounts automatically: 20% off 5–9 procedures, 35% off 10–19, 50% off 20–39 and 60% off 40 or more.
Not sure where your quality system stands? Take the free QMS gap check: about 20 questions, an instant report of your gaps by clause, and the procedures that close each one.



Comments