Toxicological Risk Assessment Under ISO 10993-17: TTC, Tolerable Intake and Margin of Safety
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.
Once a chemistry laboratory has told you which chemicals a device can release, someone has to decide whether those amounts are safe. That is the toxicological risk assessment, and ISO 10993-17:2023 sets out how to do it. Done well, it can close several biological effects without a single animal test: systemic toxicity, genotoxicity, carcinogenicity and reproductive and developmental toxicity.
This guide walks through the assessment in the order it is done, with the decisions at each step. Each summary of the standards is in our own words. The assessment itself must be carried out by a qualified toxicologist. This guide explains what they do and what you should expect to see in their report.
Who can do it
ISO 10993-17 clause 5.1.1 asks for experienced people who know toxicology, the device (clinical use, materials and manufacturing) and exposure estimation. Most device companies contract this out. The manufacturer's job is to give the toxicologist complete inputs and to check that the report contains what the standard requires.
Step 1: Assemble the inputs
Clause 5.2 starts with three inputs:
A device description: the materials that contact the body.
The intended use: indications, contraindications, use conditions, route of exposure, patient population, duration and frequency. From these, the worst-case clinical use scenario is set.
The identity and amount of each reportable constituent: usually from the chemical characterization report under ISO 10993-18, with names and CAS numbers.
The worst case drives everything. Clause 8 asks for the largest number or quantity of devices one person can be exposed to, and the lowest body mass of the people who can be exposed, with extra estimates for susceptible groups such as neonates.
Step 2: Choose a route for each constituent
Each constituent can be cleared in one of four ways:
Harm not relevant. If the harm a constituent can cause is not relevant to how the device is used, the risk is negligible.
Below the toxicological screening limit (TSL). If the total quantity of a constituent is at or below the TSL, no further assessment is needed. Amendment 1:2025 rewrote Annex B, which sets out how the TSL is calculated, and it excludes devices used for prolonged or long-term periods in preterm or very young infants. The TSL is not used where the amount cannot be measured.
Below the threshold of toxicological concern (TTC). For a constituent with no specific toxicity data, compare its worst-case daily exposure with the TTC for the contact duration.
Below the tolerable intake (TI). For a constituent with enough data, the toxicologist derives a TI (or a tolerable contact level for local effects) and calculates a margin of safety.
Step 3: Apply the TTC correctly
ISO/TS 21726:2019 sets the TTC by how long the body is exposed, and the choice must be justified:
120 µg per day for limited and prolonged contact (up to 30 days).
20 µg per day for long-term contact over 1 to 12 months.
10 µg per day for long-term contact over 1 to 10 years.
1.5 µg per day for more than 10 years up to a lifetime.
A value above the lifetime one needs use conditions that limit the maximum contact time; otherwise use 1.5 µg per day.
The TTC has firm limits. It is not used for constituents that have enough data to derive a TI, for the cohort of concern (for example N-nitroso, aflatoxin-like and azo compounds, and heavy metals), for high molecular weight polymers, particles, ceramics, proteins or radioactive constituents, for gas pathways, or for local endpoints such as cytotoxicity, irritation and sensitization. A nitrosamine from a rubber accelerator, for example, always needs a constituent-specific assessment.
Step 4: Derive a tolerable intake where needed
Clause 6.2.1 asks for a systematic search of several information sources, with the sources and search criteria documented and justified, and the reliability and quality of each study evaluated. The toxicologist picks a point of departure from the best-quality relevant data and applies uncertainty factors to reach the TI. For a human or suspected human carcinogen, clause 6.2.3 also asks for the cancer risk to be estimated or the constituent to be controlled.
Step 5: Estimate exposure and calculate the margin of safety
The worst-case estimated exposure dose (EEDmax) is the daily amount reaching the body under worst-case use, per kilogram of body mass where it is compared with a TI. The margin of safety (MoS) is simply:
MoS = TI ÷ EEDmax
Clause 9.2.2 adds one more calculation. Where constituents are released together and cause the same harm in the same target organ by the same mode of action, their margins are combined: the combined MoS is 1 divided by the sum of the reciprocals of the individual margins. Two constituents that each pass alone can fail together.
A simple example with made-up numbers: a catheter releases 200 µg of a constituent per device, up to 3 devices a day, in adults with a lowest body mass of 50 kg. EEDmax is 200 × 3 ÷ 50 = 12 µg/kg/day. With a TI of 50 µg/kg/day, the MoS is 4.2.
Step 6: Decide
Clause 10.1 sets the acceptance rule: a toxicological risk is tolerable when the MoS is above 1 and the values behind it are shown to be conservative, meaning the estimate deliberately overstates the likely true risk. For a carcinogen, the cancer risk must not exceed 1 in 100,000.
Under clause 10.2, a risk is taken further under ISO 10993-1 and ISO 14971 when the MoS is below 1, when a carcinogen's risk is above 1 in 100,000, or when expert judgement finds a possible risk. The options are:
Refine the exposure: run a leachables study under simulated use instead of relying on an exhaustive extraction.
Improve the data: find better toxicity data for a better-supported TI.
Control the risk: change the material or add a cleaning step.
Benefit-risk: a documented benefit-risk decision.
Step 7: Check the report
Clause 11 lists what the report must contain: constituent data, how the TSL or TTC were applied, how each TI or tolerable contact level was derived, the exposure estimates, each MoS and the conclusion for each constituent. If any of these is missing, send the report back.
Then carry the results into the biological evaluation. Mark the effects the assessment closed. Cytotoxicity, irritation and sensitization stay open unless you have other data, because the TTC cannot be used for local endpoints.
Existing devices
Clause 5.1.1 says marketed devices do not need historical chemical data reassessed just because of the 2023 edition. A documented justification that the earlier assessment is still adequate is enough, unless a change has occurred that triggers re-evaluation.
How the A.P. Lyon kit fits
The Biological Evaluation Kit includes the Chemical Characterization and Toxicological Risk Assessment Procedure (SOP 7-102-BIO) and an Excel Toxicological Risk Assessment Record (FORM 0106-BIO). The workbook takes the worst-case inputs once, calculates EEDmax and the margin of safety for each constituent, applies the TTC you select, flags anything that needs refinement and combines margins for same-organ groups. The procedure tells you what to give your toxicologist and what to check in the report.
Related reading: Biological Evaluation Plan and Report Under ISO 10993-1:2025 and Chemical Characterization Under ISO 10993-18.
Frequently asked questions
What margin of safety is acceptable under ISO 10993-17?
A margin of safety above 1, provided the values behind it are shown to be conservative. For carcinogens, the cancer risk must also not exceed 1 in 100,000.
When can I use the threshold of toxicological concern?
For a constituent with no specific toxicity data, outside the cohort of concern and other excluded types, and only for systemic endpoints. It cannot be used for cytotoxicity, irritation or sensitization.
Can a toxicological risk assessment replace biological testing?
For some effects, yes. ISO 10993-1:2025 allows systemic toxicity, genotoxicity and carcinogenicity to be evaluated through a toxicological risk assessment of chemical information, done to ISO 10993-17.