Biological Evaluation Plan and Report Under ISO 10993-1:2025: A Step-by-Step Route

11 hours ago
7 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


A biological evaluation is not a list of tests. Under ISO 10993-1:2025 it is a piece of risk management: you identify biological hazards, estimate the risks, compare them with criteria you set in advance, control what is not acceptable and write down the conclusion. Tests are one way to fill a gap in the data, and often not the first one.


This guide sets out the route as a sequence of stages, with who usually does each stage and what "done" looks like. Each summary of the standard is in our own words. If you are still working out which materials touch the patient and what evidence you already hold, start with our earlier guide, Medical Device Biological Evaluation: Materials, Contact and Evidence Gaps.


Who does the work


ISO 10993-1:2025 clause 4.1 asks that the evaluation be planned, carried out and reported by competent people, with evidence of their competence on file. The plan itself must name the people responsible and the parts of the work that need specialist skills, such as analytical chemistry, toxicology, biological testing and engineering.


In most device companies that means:


  • Design Assurance Engineer: owns the biological evaluation file, writes the plan and report, and manages the laboratories.

  • NPD Engineer: supplies the device description, materials and processes, and makes any design change chosen as a risk control.

  • Toxicologist: in-house or contracted, carries out the toxicological risk assessment.

  • Analytical chemistry and biological test laboratories: run the studies under approved protocols, under ISO/IEC 17025 or GLP quality controls.

  • Quality Manager: names the team, approves the plan and the report.


File the CVs, training records or accreditation certificates in the evaluation file at the start. Clause 9 asks for the report author's competence evidence again at the end.


Stage 1: Describe the device and its body contact


Start with the device, not the tests. Record the configuration, intended use, patient population, reasonably foreseeable misuse, and every process the device goes through in its life, including sterilization and, for reusable devices, the maximum number of processing cycles (clauses 4.3 and 5).


Then list every component and material: supplier, grade, colorants, coatings, adhesives, processing aids and any primary packaging that touches the device. Clause 6.2 lists the characteristics to consider, including the effect of manufacturing steps such as laser marking, passivation or sterilization, and constituents added during manufacture or reprocessing.


Categorize each contacting component separately (clause 6.4). The contact types are intact skin; intact mucosal membrane; breached or compromised surfaces or internal tissues other than circulating blood; and circulating blood. Duration is limited (up to 24 hours), prolonged (more than 24 hours up to 30 days) or long-term (more than 30 days), counted as the total exposure on one patient. For daily use, count calendar days from first to last use; for intermittent use, add the contact days. If more than one category could apply, use the longest and record why. A component with no direct or indirect contact needs no biological evaluation (clause 6.4.4), but write down why it is out.


Done when: every contacting component has a contact type and a duration category.


Stage 2: Gather what is already known


Clause 6.1 asks you to use all relevant information: supplier composition data, clinical use history of the same materials, earlier test reports, literature, complaints and adverse event data. Where a source is incomplete or weak, justify its quality and relevance in writing.


Earlier test reports need a closer look. Clause 6.8.1 expects testing on the finished device (or a justified representative sample) after sterilization, by a laboratory under appropriate quality controls, and asks you to justify any difference between how an old test was run and the current standard.


This is also the point to decide whether to claim biological equivalence to a device with existing safety data (clause 6.7). The comparison covers type and duration of contact, composition, relevant physical properties, manufacturing processes including packaging and sterilization, patient population, and size or quantity of material. Equivalence holds only where no difference creates a new or increased biological risk, and the claim must be documented step by step: chemical, physical, material, contact and finally biological equivalence. If it holds, biological safety is established for the new device and you can go straight to the report.


Done when: each data source is reviewed and the equivalence decision is recorded.


Stage 3: Write the biological evaluation plan


Clause 5 requires a documented plan. It includes or refers to the device description, intended use and misuse, categorization, the characteristics related to biological safety, the biological hazards and harms, the acceptability criteria, the gap analysis, the strategy and planned content of the report, the criteria for the overall conclusion, the specialist skills needed and the responsible people. The plan may start as an outline and grow as data come in.


The working part of the plan is an effect-by-effect table:


  1. For each contacting component, list the biological effects shown for its category in Tables 1 to 4 of ISO 10993-1. Clause 6.5.1 is clear that these tables list effects to address, not tests to run.

  2. Add effects that apply whatever the category: a known hazardous constituent, degradation or wear, particulates, toxicokinetic triggers, contact with nervous tissue, reproductive toxicity triggers, immunotoxicity and material-mediated pyrogenicity (clauses 6.5.10 and 6.5.11).

  3. For each effect, decide how it will be addressed: existing data, chemistry plus toxicological risk assessment, a biological test, or not applicable with a justification. Very brief contact and low-risk intact skin devices can rest on a written justification (clauses 6.5.11.2 and 6.5.11.3).

  4. Do the gap analysis (clause 6.6.1): is the information enough to estimate each risk? If not, plan the data to generate.

  5. Set the acceptability criterion for each effect, for example a margin of safety above 1 in the toxicological risk assessment, or the test-specific criteria of ISO 10993-5 for cytotoxicity.


Plan chemistry before animal tests. Clause 4.4 puts animal welfare into the requirements: no in vivo test where valid data already address the effect, and none where released constituents are already characterized and known to be acceptable.


Done when: the plan is approved, before new testing starts.


Stage 4: Generate the data


Chemical characterization comes first (clause 6.3 requires it at least for the materials of construction and the processing aids and additives, done to ISO 10993-18). The toxicological risk assessment to ISO 10993-17 then closes the systemic effects it can: systemic toxicity, genotoxicity, carcinogenicity and reproductive toxicity. We cover both in Chemical Characterization Under ISO 10993-18 and Toxicological Risk Assessment Under ISO 10993-17.


Cytotoxicity, irritation and sensitization cannot be closed with the threshold of toxicological concern, so these usually need data of their own. ISO 10993-1 asks for all three on every body-contacting device. ISO 10993-23 puts the in vitro reconstructed human epidermis test before any animal irritation test, and ISO 10993-10 makes the LLNA the preferred in vivo sensitization assay where a test is still needed.


When a test fails, it is not an automatic rejection. ISO 10993-5 clause 10 asks for a competent person to assess a cytotoxic result alongside the other data and the intended use.


Stage 5: Estimate, evaluate and control the risk


For each biologically hazardous situation, meaning a hazard with real exposure, estimate severity and probability with the method in your risk management procedure (clause 6.9). Where data are limited, use reasonable worst-case assumptions and say so.


Then compare each risk with its criterion (clause 7). If it is met, biological safety is established for that effect. If not, you have two routes: refine the estimate (better quantification, a better-supported tolerable intake, or a test) or control the risk. Clause 8 sets the order of controls:


  1. Inherently safe design and manufacture, for example a less toxic material.

  2. Protective measures in the device or the process, for example a barrier layer or an added cleaning step.

  3. Information for safety and training, for example a contraindication or warning.


A risk still not acceptable after control as far as feasible goes to a benefit-risk assessment under ISO 14971, and significant residual biological risks must be disclosed to users.


Stage 6: Write and approve the report


Clause 9 lists what the biological evaluation report records: the acceptability criteria, conformity with the plan, how each applicable effect was addressed, interpretation of all data, the reasons for tests run or waived and for any non-standard method, the risk evaluation results, risk controls, residual risks and the overall safety conclusion. Competent people prepare it, and their competence evidence goes with it.


Stage 7: Keep it current


Clause 10 asks for a system that collects information relevant to biological safety in production and post-production. Review and update the evaluation when post-market information shows a new hazard, or when any of these change: the specification, formulation, processing or source of a contacting material; processing, primary packaging or sterilization; storage or shelf life; or intended use. Each review is documented, with an overall safety statement and the scientific reason for doing, not doing or limiting further work.


For devices already on the market, clause 6.6.2 says the 2025 edition is not a reason to retest by itself. Review the historical information, record whether any difference (design or use changes, new effects, literature, adverse events) affects the conclusion, and justify the outcome.


How the A.P. Lyon kit fits


The Biological Evaluation Kit turns these stages into 4 procedures and 12 forms written as numbered steps, each with an owner, the record to fill in and the clause reference in a separate column. It includes the device and contact categorization form, the existing data review, the biological evaluation plan and report, an Excel toxicological risk assessment calculator, sample preparation, test request and report review forms, and change and periodic review forms. It works alongside the Biocompatibility and Toxicity Testing Procedure in our Manufacturing Plus systems.


Frequently asked questions


Is a biological evaluation the same as biocompatibility testing?


No. Testing is one source of data. ISO 10993-1:2025 asks for a planned evaluation within risk management, with criteria set in the plan, and many effects can be addressed by existing data or by chemistry and toxicological risk assessment.


Do I have to retest marketed devices for ISO 10993-1:2025?


Not because of the new edition alone. Clause 6.6.2 asks you to review the existing evaluation and document whether any change or new information affects the conclusion.


What must the biological evaluation plan contain?


Clause 5 lists the device description, intended use and misuse, categorization, safety-related characteristics, hazards, acceptability criteria, gap analysis, report strategy, overall conclusion criteria, specialist skills and responsible people.