Chemical Characterization Under ISO 10993-18: Extraction, AET and What to Ask Your Lab
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.
ISO 10993-1:2025 makes chemistry the first source of data in a biological evaluation. Clause 6.3 requires chemical characterization, done to ISO 10993-18, at least for the materials of construction and the processing aids or additives used in manufacture. The results feed the toxicological risk assessment, which can close systemic effects without animal tests.
Most companies send this work to a contract laboratory. The quality of the result depends on what you ask for. This guide covers the decisions you make before the samples leave your building, and what to check when the report comes back. Each summary of the standards is in our own words.
Step 1: Start with what you already know
ISO 10993-18 is step by step. You gather existing information first and generate new data only when it is not enough (clause 5.1). Clause 5.2 asks you to document the device configuration, intended use, materials of construction and their proportions, with each material's identity: trade name or specification, supplier and composition data.
Ask your suppliers for formulation data, including additives, residual monomers and processing aids. Remember the things that are not on the drawing:
cleaning and disinfection agents;
sterilant residues;
mould release agents;
anything the primary packaging can transfer.
Note how processing changes the material. Sterilization is the obvious case.
Step 2: Try the hypothetical worst case
If you know the full composition, clause 5.4 lets you assume every bit of each constituent is released and hand that to the toxicologist. If the risk is acceptable even on that basis, chemical characterization is complete and no extraction study is needed. This is often the case for simple devices made from well-documented materials.
If you start the process at a later point, for example going straight to an extraction study, write down why.
Step 3: Choose the extraction approach
When an extraction study is needed, the approach depends on how long the device contacts the body. Table 2 of ISO 10993-18 recommends:
Limited contact (up to 24 hours): simulated use conditions, with exaggerated conditions as a credible alternative. Some regulators, for example FDA, may ask for exaggerated extraction.
Prolonged contact (more than 24 hours to 30 days): exhaustive conditions, with exaggerated as an alternative.
Long-term contact (more than 30 days): exhaustive conditions, with exaggerated as an alternative.
Exaggerated extraction can be enough in some prolonged or long-term cases, for example a single-use device replaced daily or a reusable device, where the extraction accounts for each individual use. Whatever you choose, clause 5.6 asks you to document and justify it. One extraction per vehicle is enough.
Step 4: Specify the test article and conditions
Give the laboratory a written protocol, not just samples. Specify:
The test article: the finished, sterilized device or a justified representative portion. ISO 10993-1 clause 6.8.1 asks for the finished device or material processed the same way.
The number of devices to extract.
Extraction vehicles: polar and non-polar, with others where justified.
Conditions: temperature, duration, cycles, the surface-area or mass ratio, and agitation.
Step 5: Set the analytical evaluation threshold
The AET is the concentration in the extract above which the laboratory must identify and report a chemical. It is calculated from a dose-based threshold, often the TTC for the contact duration from ISO/TS 21726, together with the number of devices extracted and the extract volume. Clause 5.5 asks for the AET to be set and justified.
Amendment 1:2022 added an important adjustment. Different chemicals give different instrument responses, so the AET is lowered by an uncertainty factor for that variation, and the value and how it was set are justified. If the variation is so large that the adjusted AET falls below what the method can detect, the AET approach is not used. Instead, every observed peak is identified and quantified, or the method is improved.
Name the analytical methods and how each will be shown suitable. Typical choices are GC-MS for volatile and semi-volatile organics, LC-MS for non-volatile organics and ICP-MS for elements. Methods used to quantify leachables must be qualified for that use.
Step 6: Check the report
Clause 7 lists what the chemical characterization report must describe and justify:
the test article and sample preparation;
the analytical methods and extraction conditions (vehicles, duration and cycles, temperature, ratio, agitation);
system suitability testing and its outcome;
the reporting threshold (AET) and its justification;
the identities of extractables and how they were identified;
their concentrations and how they were quantified.
If any item is missing, ask for it before the report goes to the toxicologist.
Then list every constituent at or above the AET with its name, CAS number, amount per device and confidence of identification. Flag any cohort-of-concern constituent, such as a nitrosamine, and any known sensitizer, carcinogen or reproductive toxicant. ISO 10993-1 clause 6.5.11.9 requires a constituent of known toxicity at potentially toxic levels to be addressed whatever the device category.
Step 7: Refine with leachables if needed
If the toxicological risk assessment cannot accept a constituent on the extractables data, the next step is often a leachables study under simulated use (ISO 10993-18 clause 5.8). It shows what is actually released in clinical use rather than the worst case. Methods used to quantify leachables must be qualified for that use.
How the A.P. Lyon kit fits
The Biological Evaluation Kit includes the Chemical Characterization and Toxicological Risk Assessment Procedure (SOP 7-102-BIO), a Chemical Characterization Study Protocol (FORM 0104-BIO) that takes you through the extraction choice, the AET and the method selection, and a Results Review (FORM 0105-BIO) with the report content checklist and a constituent list ready for the toxicologist.
Related reading: Toxicological Risk Assessment Under ISO 10993-17 and Biological Evaluation Plan and Report Under ISO 10993-1:2025.
Frequently asked questions
Do I always need an extractables study?
No. If the full composition is known and the toxicologist finds the hypothetical worst-case release acceptable, chemical characterization is complete without one (ISO 10993-18 clause 5.4).
Exhaustive or exaggerated extraction?
ISO 10993-18 Table 2 recommends exhaustive extraction for prolonged and long-term contact and simulated use for limited contact, with exaggerated conditions as a credible alternative. Document and justify your choice.
What is the AET uncertainty factor?
Amendment 1:2022 lowers the AET by an uncertainty factor to allow for differences in instrument response between chemicals, so weakly responding chemicals are not missed. The value and how it was set must be justified.