ISO 14644-1 Sampling Locations and Sample Volume: How to Plan a Cleanroom Classification Test

6 hours ago
7 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


A cleanroom classification test is only as good as its plan. If the room has too few sampling locations, or each sample is too small to show a count near the class limit, the result says very little, and an auditor who checks the numbers will find it. ISO 14644-1:2015 sets out how to work out both. This guide walks through the plan in the order a team does it: state the class, count the locations, size the samples, run the test, judge the results and write the report.


Every summary of the standard here is in our own words. ISO 14644-1 is a copyrighted standard; buy a licensed copy before you test. For the wider picture of cleanroom procedures in a device quality system, see Medical Device Cleanrooms: Qualification, Monitoring, Cleaning and Gowning.


Step 1: State the class designation


Before anything is measured, write down exactly what the room is being tested against. A designation has three parts:


  • the ISO Class number, for example ISO Class 7

  • the occupancy state: as-built, at-rest or operational

  • the considered particle size or sizes, for example 0.5 µm and 5 µm


Written out, that reads "ISO Class 7; at rest; 0.5 µm, 5 µm". Particle sizes are chosen between 0.1 µm and 5 µm. Where more than one size is used, each larger size must be at least 1.5 times the next smaller one, so 0.5 µm and 5 µm is fine, but 0.5 µm and 0.6 µm is not.


The class limit for each size is the maximum number of particles per cubic metre at or above that size. ISO 14644-1 gives those limits in Table 1 and in Formula E.1, Cn = 10^N x (0.1 / D)^2.08, where N is the class number and D is the particle size in micrometres. Some cells of Table 1 have no limit and cannot be used. ISO Class 5 at 5 µm is one of them; where 5 µm particles matter at Class 5, the standard's macroparticle (M) descriptor is used instead. ISO Class 9 applies only in operation.


Two limits come up often in device assembly rooms:


  • ISO Class 7: 352,000 per m³ at 0.5 µm and 2,930 per m³ at 5 µm

  • ISO Class 8: 3,520,000 per m³ at 0.5 µm and 29,300 per m³ at 5 µm


Step 2: Work out the number of sampling locations


The minimum number of sampling locations, NL, comes from Table A.1 of the standard, which is based on the area of the room or zone. Use the horizontal plan area. For unidirectional airflow, the area may instead be the cross-section of the moving air perpendicular to the flow.


When the area falls between two rows of Table A.1, use the row for the next larger area. Some examples:


  • a 10 m² room needs at least 5 locations

  • a 30 m² room falls between the 28 and 32 m² rows, so the 32 m² row applies: at least 8 locations

  • a 100 m² room uses the 104 m² row: at least 16 locations

  • above 1,000 m², NL = 27 x (area / 1,000), rounded up, so a 1,500 m² hall needs at least 41 locations


Then place them:


  1. Divide the area into NL sections of equal area.

  2. Choose one representative location in each section.

  3. Put the probe in the plane of the work activity, or at another point you have specified.

  4. Add any extra locations for critical points, such as a pouch sealer or a pass-through, and record that they were agreed.


In rooms with non-unidirectional airflow, a location directly beneath a non-diffused supply air outlet may not be representative of the room. Move it within its section and record why.


One change from the older edition is worth knowing: ISO 14644-1:2015 no longer uses the 95% upper confidence limit calculation. Each location must meet the class limit on its own.


Step 3: Size the samples


Each sample must be big enough that 20 particles would be counted if the air were exactly at the class limit for the largest considered size. The minimum single sample volume is:


Vs = (20 / Cn,m) x 1,000 litres


where Cn,m is the class limit for the largest considered particle size. Two more minimums apply: each sample is at least 2 litres and at least 1 minute long. The single sample volume must be the same at every location.


Worked through for the designations above:


  • ISO Class 7 at 0.5 µm and 5 µm: the largest size is 5 µm, with a limit of 2,930. Vs = 20 / 2,930 x 1,000 = 6.8 litres. A 28.3 litre per minute counter running for 1 minute takes 28.3 litres, which meets all three minimums.

  • ISO Class 8 at 0.5 µm and 5 µm: the 5 µm limit is 29,300, so Vs works out at 0.68 litres. The 2 litre and 1 minute minimums then govern.

  • ISO Class 5 at 0.5 µm only: the limit is 3,520, so Vs = 5.7 litres.


Where Vs is very large, for example in the cleanest classes, the sequential sampling procedure in Annex D of the standard may be used.


Step 4: Set up and run the test


Before testing, the installation must be complete and working to its performance specification. Then:


  • check that the particle counter has a valid calibration certificate. Calibration frequency and method follow current accepted practice in ISO 21501-4. If the counter cannot be calibrated to every test in ISO 21501-4, record the decision to use it in the test report.

  • set the counter up to the manufacturer's instructions, including a zero count check

  • point the probe into the airflow, or straight up where the airflow direction is not controlled

  • establish normal conditions for the occupancy state, for example a settling time for at-rest testing, before the first sample


Two kinds of high count are handled differently:


  • An identified abnormal occurrence, such as a door propped open during the sample. The count may be discarded, noted in the report and the location resampled.

  • A technical failure of the cleanroom or the equipment. The cause is found and fixed, then the failed location, the locations around it and any others affected are retested. Record the choice of locations and the reason.


Step 5: Judge the results


Record the count of each sample at each considered size. Where a location has two or more samples, average them. The concentration per cubic metre is the average count x 1,000 / the sample volume in litres.


The room or zone meets the class when the average concentration at every location is at or below the class limit for every considered size. All sizes are measured by the same method. Any out-of-specification count is investigated, and the result and the remedial action go in the test report.


Step 6: Write the test report


The report ends with a statement that the room or zone does or does not comply with the specified class. It also records:


  • the name and address of the testing organization, and the test date

  • the standard used: ISO 14644-1:2015

  • where the room or zone is, and the coordinates of every sampling location (a diagram helps)

  • the designation: class, occupancy state and particle sizes

  • the test method, any special conditions or departures, and the instrument with its current calibration certificate

  • the particle data for every location, and macroparticle results if they were measured


Common mistakes


  • Using the room's old location count after the room was enlarged or split. Recalculate from Table A.1 every time the area changes.

  • A sample volume worked out from the wrong size. It is the largest considered size that sets Vs, not the smallest.

  • Different sample volumes at different locations. The single sample volume is the same everywhere.

  • Discarding a high count without a reason. Only an identified abnormal occurrence allows it, and the report must say so.

  • A report with no location coordinates or calibration record. Both are required, and both are among the first things an auditor asks for.


Putting it into practice


The Cleanroom Qualification and Monitoring Kit includes an Excel Sampling Location and Sample Volume Calculator. Enter the area, class, occupancy state and particle sizes, and it gives the minimum number of locations, the class limits and the sample volume and time, then judges up to 60 locations as COMPLIES or DOES NOT COMPLY. It is used under a step-by-step Cleanroom Classification Test Procedure, with a test report form that holds every item listed above.


Frequently asked questions


Can I test fewer locations than Table A.1 gives?


No. Table A.1 gives the minimum. You can add locations for critical points, but not remove them.


How often must a cleanroom be classified?


ISO 14644-2:2015 says annually. The interval may be extended only on the basis of a risk assessment, the extent of monitoring and data that stay consistently within limits. See Cleanroom Excursions and Requalification.


Does a classification test replace routine monitoring?


No. Classification shows the room meets its class on the test day. Routine monitoring under a written plan shows it keeps doing so. See How to Write an ISO 14644-2 Cleanroom Monitoring Plan.