ISO 15223-1 Symbols on Medical Device Labels: What Goes Next to Each Symbol and When the FDA Needs a Symbols Glossary

7 hours ago
5 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


Symbols save label space and work across languages, which is why most device labels use them. But a symbol on its own is not always enough. Some must carry a name, a code or a date next to them. In the US, a symbol used without words beside it needs a symbols glossary and a statement on the labeling saying where to find it.


This guide covers the rules in ISO 15223-1:2021 with Amendment 1:2025, and the FDA rules in 21 CFR 801.15. Every summary of a requirement is in our own words; the standard itself is copyrighted, so buy a licensed copy for the symbol graphics and full text.


When symbols may be used


ISO 15223-1 clause 4.2 says that where a manufacturer decides symbols are the right way to convey information essential for proper use, the symbols in its Table 1 may be put on the device, its packaging or the accompanying information. Clause 5 adds that a manufacturer may use any appropriate symbol, and clause 4.3 notes that other standards contain symbols for particular devices.


In the EU, MDR Annex I 23.1(h) asks that symbols and identification colours conform to harmonised standards or common specifications, or, where there are none, be described in the documentation supplied with the device.


Symbols that need information next to them


The Requirements column of Table 1 is where most errors are found. These symbols must be accompanied by information adjacent to the symbol:


  • Manufacturer (5.1.1): the manufacturer's name and address.

  • Authorized representative (5.1.2): the representative's name and address. Amendment 1:2025 made this symbol generic: the text in the symbol is replaced by the country code or other text the authority requires.

  • Date of manufacture (5.1.3) and use-by date (5.1.4): the date, expressed in line with ISO 8601-1. The use-by date means the device should not be used after the end of the year, month or day shown.

  • Batch code (5.1.5), catalogue number (5.1.6), serial number (5.1.7) and model number (5.1.10): the code or number.

  • Importer (5.1.8) and distributor (5.1.9): the entity's details.

  • Patient information website (5.7.4): the web address.


When one entity is, for example, both importer and distributor, the name and address need not be repeated; the symbols can be grouped next to one address.


Sterile and sterile barrier symbols


The 5.2 group covers the sterilization method symbols and related symbols such as do not resterilize, non-sterile, do not use if the package is damaged, and sterile fluid path. The single and double sterile barrier system symbols (5.2.11 to 5.2.14) are placed adjacent to, or combined with, the sterilization method symbol. Check that the method shown matches the method actually used.


The UDI symbol


The UDI symbol (5.7.10) is optional. It is most useful when a label carries more than one bar code. If it is used, it is placed next to the UDI carrier.


Size and dates


Clause 4.2 b) leaves the size of each symbol to the manufacturer, who sets it so it is legible for its purpose. Clause 4.2 c) asks for dates and times shown with symbols to follow ISO 8601-1 and ISO 8601-2. In the US, 801.18 separately requires every date on a device label to be written as YYYY-MM-DD.


The FDA rules for symbols without words


21 CFR 801.15(c) allows a symbol to be used without adjacent explanatory text only if one of two conditions is met:


  1. The symbol is in a standard FDA recognizes and is used as FDA's recognition specifies, or

  2. The symbol is in a standard from a standards development organization, and the manufacturer has determined it is likely to be read and understood by the ordinary user under customary conditions.


Either way, the symbol must be explained in a paper or electronic symbols glossary included in the labeling, and the labeling on or within the package must carry a prominent statement saying where the glossary is. A symbol used without words that does not meet these rules misbrands the device.


The glossary lists, for each symbol:


  • the symbol

  • the title and number of the standard it comes from

  • the symbol's title and reference number

  • its meaning


Many manufacturers place the glossary at the back of the IFU and add a short line on the pouch or box pointing to it.


A symbol record keeps it straight


For each label set, a symbol record lists every symbol chosen, the standard and edition, its reference number, the information placed next to it and where it is used. The same record holds the basis for each symbol used in the US without words, and the glossary text. When a new edition of ISO 15223-1 comes out, the record shows at once which labels are affected.


Putting it into practice


The Labeling and UDI Kit includes a Symbol Selection Record and Symbols Glossary form, used under a step-by-step Label Content and Symbol Selection Procedure, and a Label Artwork Approval form that checks symbols, adjacent information and the glossary statement on every proof. For the full label content by market, see Medical Device Label Content for the US and the EU.


Frequently asked questions


Can I use a symbol that is not in ISO 15223-1?


Yes. Clause 5 allows any appropriate symbol, and other standards contain symbols for particular devices. In the US the 801.15(c) conditions and the glossary still apply.


Do I still need the glossary if every symbol has words next to it?


801.15(c)(1)(i)(C) allows a symbol accompanied by adjacent explanatory English text. The glossary rules apply to symbols used without that text.


Did Amendment 1:2025 change the authorized representative symbol?


Yes. It made the symbol generic, with its text replaced by the country code or other text the authority requires, and accompanied by the representative's name and address.


Building a complete quality system? Build Your QMS lets you choose ISO 13485 + FDA QMSR, Dual-Standard or ISO 9001:2026 procedures in one place, and applies volume discounts automatically: 20% off 5–9 procedures, 35% off 10–19, 50% off 20–39 and 60% off 40 or more.


Not sure where your quality system stands? Take the free QMS gap check: about 20 questions, an instant report of your gaps by clause, and the procedures that close each one.


Just getting started? The free Startup Quality Plan shows which procedures you need now, which can wait, and how long they take to put in place.