Medical Device Label Content for the US and the EU: Building a Label Content Matrix From 21 CFR 801 and MDR Annex I Section 23

7 hours ago
6 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


Most labeling mistakes start before anyone draws a label. A required element is missed because no one listed what each market asks for, or a statement is written for one market and copied to another where it does not fit. A label content matrix fixes that. It lists every element each target market requires on the label, the sterile packaging and the instructions for use, and records where each one goes and what it says.


This guide shows how to build one for the US (21 CFR Part 801) and the EU (Regulation (EU) 2017/745, Annex I Section 23). Every summary of a requirement is in our own words. For the inspection and release side of labeling, see Labeling and Packaging Controls Under the QMSR.


Start with the device facts


The elements that apply depend on a short list of facts. Record them first:


  • the device name, models and sizes, intended use and users

  • prescription or over-the-counter (in the US)

  • sterile (and the method) or non-sterile

  • single use, single patient multiple use, or reusable and reprocessed

  • whether the device or its packaging contains natural rubber latex that contacts humans

  • whether it contains a medicinal substance or tissues or cells of human or animal origin

  • implant or active implant; stand-alone software

  • each packaging level (unit, box, case) and what it contains

  • the target markets and the languages each one requires


Each fact switches rows on or off. A non-sterile device does not need the sterile pack rows. A prescription device in the US needs the Rx only rows.


The US elements under 21 CFR 801


The main US label requirements in Part 801 are:


  • Name and place of business (801.1): the label of a packaged device states the name and place of business of the manufacturer, packer or distributor, using the actual corporate name. If the named firm did not make the device, the name is qualified, for example "Manufactured for" or "Distributed by". The address includes street, city, state and ZIP code.

  • Directions for use (801.5): labeling gives adequate directions for use, unless the device is a prescription device that meets 801.109.

  • Prescription devices (801.109): the label bears "Rx only", "℞ only" or the caution statement restricting sale to the order of a licensed practitioner, plus the method of use. Labeling on or within the dispensing package gives information for use, including indications, hazards, contraindications, side effects and precautions. All labeling other than labels and cartons bears its issue or latest revision date.

  • Prominence and language (801.15): required statements are prominent and conspicuous, in English, and if any foreign language appears, all required information also appears in that language.

  • Dates (801.18): every date on the label is written as YYYY-MM-DD.

  • UDI (801.20 and 801.40): the label and every device package bear a UDI in plain text and in a machine-readable form, unless an exception applies.

  • Latex (801.437): devices with natural rubber that contacts humans carry the required caution statements.


Over-the-counter devices also need a statement of identity and a net quantity declaration on the principal display panel (801.61 and 801.62).


The EU elements under MDR Annex I Section 23


The EU rules are longer and more prescriptive. Section 23.2 lists the label particulars, from (a) to (s). The ones that apply to most devices are:


  • the name or trade name, and the details a user needs to identify the device and the package contents

  • the manufacturer's name and address, and the authorized representative's name and address if the manufacturer is outside the EU

  • the lot or serial number, with the word or an equivalent symbol

  • the UDI carrier

  • the use-by date, at least as year and month, or the date of manufacture where there is no use-by date

  • special storage and handling conditions

  • the sterile state and sterilization method, where the device is sterile

  • warnings or precautions that need the user's immediate attention

  • a single-use indication, where applicable

  • an indication that the product is a medical device


Section 23.3 lists what goes on sterile packaging, including the month and year of manufacture and an instruction to check the instructions for use if the packaging is damaged or opened before use. Section 23.4 lists the contents of the instructions for use, including the intended purpose, indications, contra-indications, target groups, performance characteristics, residual risks and side-effects, and the date of issue or latest revision.


Article 10(11) adds that the information must be in the official Union language or languages set by the Member State where the device is made available.


Build the matrix


A workable matrix has one row per element and these columns:


  1. where the element belongs (label, sterile pack, IFU, device marking, software)

  2. the element

  3. the US rule and a Y, N or V (voluntary) mark for the US

  4. the EU rule and a Y, N or V mark for the EU

  5. other markets, if any

  6. the location on the artwork (unit, box, sterile pack, IFU)

  7. how it is shown (text, symbol or both)

  8. the exact wording or the data source, for example "lot number from the work order"

  9. the symbol reference, if a symbol is used


Every row marked Y in at least one market needs a location and wording before the artwork is drawn. A simple status column that flags rows still open saves an artwork revision later.


Take warnings from the risk file


Residual risks that must be told to the user are part of the label content. ISO 14971 asks for residual risks to be disclosed in the accompanying documentation, and MDR Annex I 23.1(g) asks that they appear as limitations, contra-indications, precautions or warnings. Pull them from the risk management file into the matrix, and check each one has a place on the label or in the IFU.


Common gaps


  • US dates in another format. 801.18 requires YYYY-MM-DD for every date intended for the user.

  • A missing glossary statement. Symbols used without adjacent text in the US need a symbols glossary and a statement on the labeling saying where it is (801.15(c)).

  • No revision date on the IFU. Both 801.109(e) and MDR Annex I 23.4 ask for it.

  • The EU authorized representative left off a device made outside the EU.

  • One label set for every market without checking the language rules of each.


Putting it into practice


The Labeling and UDI Kit includes an Excel Label Content Requirements Matrix with 37 US and EU label, sterile pack and IFU elements already listed, a status flag for each row and a summary that shows when the labeling is ready for artwork. It is used under a step-by-step Label Content and Symbol Selection Procedure, with the symbol record and glossary on the next form.


Frequently asked questions


Does the matrix replace the labeling procedure?


No. The labeling procedure sets the company's rules for review and approval. The matrix is the record that shows what each label must contain and why.


What about markets other than the US and the EU?


Add rows for them from the official text of each market's rules, with the rule reference in the "other market" column.


Who should own the matrix?


Usually the person who prepares the artwork fills it in, and Regulatory Affairs marks which elements each market requires. Quality Assurance approves it before artwork starts.