Labeling and Packaging Controls Under the QMSR: What 21 CFR 820.45 Requires
Updated: 5 hours ago
On 2 February 2026, FDA's Quality Management System Regulation (QMSR) took effect and incorporated ISO 13485:2016 by reference. Most of the old Quality System Regulation was replaced by the ISO clauses, but FDA kept a short list of supplemental provisions where it wanted more detail than the standard gives. Labeling and packaging is one of them. 21 CFR 820.45, Device labeling and packaging controls, sits on top of ISO 13485 clause 7.5.1 and spells out what you must inspect, what you must record and how you must prevent mixups.
If your labeling procedure still cites the old 820.120 and 820.130 sections, or treats label control as a one-time artwork approval, this is the section to read closely. Here is what it says and how to turn it into procedures that hold up in an inspection.
What 21 CFR 820.45 requires
The opening paragraph requires each manufacturer to document and maintain procedures that cover the integrity, inspection, storage and operations for labeling and packaging. Those procedures have to work under the customary conditions of processing, storage, handling and distribution and, where appropriate, use. The section then adds three specific requirements.
(a) Examine labeling and packaging for accuracy
Labeling and packaging must be examined for accuracy before release or storage, where applicable. The regulation lists what the examination must cover:
The correct unique device identifier (UDI) or universal product code (UPC), or other device identification
Expiration date
Storage instructions
Handling instructions
Any additional processing instructions
(b) Document the release of labeling
The release of labeling for use must be documented in accordance with ISO 13485 clause 4.2.5, Control of records. A signed artwork proof is not enough on its own. You need a record showing that a specific labeling revision was released for use.
(c) Prevent mixups during operations
Labeling and packaging operations must be established and maintained to prevent mixups. That includes inspection before use to confirm that the correct labeling and packaging have been selected, as specified in the medical device file. The results of that inspection must be documented under clause 4.2.5.
The UDI records that go with it
Section 820.45 does not stand alone. Other parts of the QMSR tie labeling to device identification records:
820.10(b)(1) requires a UDI system that meets 21 CFR Part 830, as part of ISO 13485 clause 7.5.8, Identification.
820.35(c) requires the UDI to be recorded for each medical device or batch.
820.35(a)(3) and (b)(2) require complaint and servicing records to include any UDI or UPC and other device identification.
In practice, the UDI you verify on the line becomes the identifier your complaint, servicing and traceability records depend on later. An error at labeling travels through the whole post-market system.
Turning the regulation into procedures
A workable approach splits the requirements into controls that each have an owner and a record:
Label content and artwork review. Define label and IFU content from design outputs and regulatory requirements, then review and approve each revision under document control.
Labeling release. Release an approved revision for use and record the release, so production only issues current labeling.
Label storage. Store labels in a controlled area, separated by product and revision, with obsolete stock removed and destroyed.
Line clearance. Clear the line of previous product, labels and packaging before each run and record it.
Inspection before use. Check the selected labels and packaging against the medical device file before the run starts, covering each item listed in 820.45(a).
Packaging integrity. Show that the package protects the device under customary storage, handling and distribution conditions. For sterile-barrier packaging this usually means qualified sealing equipment and a validated process.
Label issuance and reconciliation (counting labels issued, used, damaged and returned) are not named in 820.45, but many manufacturers use them as the most direct way to show that mixups are prevented.
Common weak spots
When teams review their labeling procedures against 820.45, the same gaps tend to appear:
The procedure covers artwork approval but says nothing about inspection on the line.
The line check confirms the part number but not the UDI, expiration date, storage or handling instructions.
There is no record that a labeling revision was released for use, only that it was approved.
Obsolete labels remain in the label room or at the packaging station.
Line clearance is done but not recorded, or the form does not cover labels and packaging materials.
The procedure still cites 820.120, 820.130 and the device master record instead of 820.45 and the medical device file.
A short self-check
Pick one recent production lot and try to answer these questions from records alone:
Which labeling revision was used, and where is the record of its release?
Who inspected the labels and packaging before use, and what did they check against?
Does the inspection record cover every item in 820.45(a) that applies to the device?
Was the line cleared before the run, and is that recorded?
Is the UDI recorded for the device or batch?
If any answer depends on someone's memory, the procedure or the forms need work. For the wider QMSR update, see QSR to QMSR: What to Change in Your Procedures.
Documents that cover these controls
A.P. Lyon documents are editable Word and Excel files that work on paper or in an eQMS. These cover the labeling and packaging controls described above:
Medical Device Labeling Procedure Bundle: four editable procedures covering label review and approval, UDI, product stability and shelf life, and CE marking.
ISO 13485 Label Review and Approval Procedure: forms for unit, box and shipper label checks, IFU content review, approvals and controlled labeling revisions.
Unique Device Identification Procedure: GUDID data entry, issuing-agency selection, packaging levels and change control, with references to 21 CFR Part 830 and EU MDR.
Line Clearance Procedure: line-clearance forms for manufacturing, label-printing and packaging areas.
For sterile-barrier packaging, the Tray Sealer Qualification Protocol Bundle and related sealer protocols are in the ISO 13485 QMS category.
If you need several procedures, Build Your QMS applies 20% off 5–9 procedures, 35% off 10–19, 50% off 20–39 and 60% off 40 or more. If you would rather have help applying 820.45 to your own lines, A.P. Lyon founder Aaron Lyon, a former Director of Quality Assurance and International Regulatory with more than 30 years in product design, development and global manufacturing, offers consulting.
Need the forms and a step-by-step route? The Labeling and UDI Kit turns this into numbered steps, each with an owner and a record. For the full route, see Medical Device Label Content for the US and the EU.



Comments