Implant Residue Limits and Exhaustive Extraction Under ISO 19227: TOC, THC, Endotoxin and Inorganics

10 hours ago
6 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


A cleaning validation is only as good as its acceptance criteria and the tests behind them. A limit with no basis, or a test that never pulled the residue off the implant in the first place, gives a passing result that proves nothing. ISO 19227:2018 sets out which tests to consider, a few limits of its own, and how extraction and test methods must be shown to work.


This guide covers those requirements in our own words. For the full validation route, see Implant Cleaning Validation Under ISO 19227: A Step-by-Step Route.


The tests to consider


Clause 5.1 lists seven tests to consider for each critical in-process cleaning and the final cleaning:


  • visual inspection;

  • bioburden;

  • bacterial endotoxins;

  • organic contaminants;

  • inorganic contaminants;

  • particulate contamination; and

  • cytotoxicity.


A test may be left out if adequately justified, based on the implant's cleanliness requirements, the characteristics of production and the cleaning process, data from earlier cleaning processes and historical data. Other tests can be added for other contaminant types. Record each decision with its reason.


Preliminary first, final later


Clause 5.1 says the preliminary cleanliness requirements are set by documenting the acceptance criteria for each test. The final requirements are set only after the results of the biological evaluation under ISO 10993-1 and the sterilization validation are available. Plan for the limits to be reviewed once those results come in.


Visual inspection


The implant manufacturer sets the acceptance criteria for visible contaminants remaining after cleaning, and inspected implants must meet them (clause 5.2). The standard names EN 13018 as one method for visual inspection criteria. Define the lighting, magnification and viewing conditions so that two inspectors reach the same answer.


Bioburden


If a purpose of the cleaning is to keep bioburden at or below a predetermined level, clause 5.3 says bioburden is determined by ISO 11737-1 on specimens that have not been sterilized. The level must be low enough for the specified sterilization method to reach the required sterility assurance level, so agree it with whoever owns the sterilization validation. Our guide to bioburden and sterility testing covers the ISO 11737-1 method.


Bacterial endotoxins


If a purpose of the cleaning is to reduce endotoxin, clause 5.4 asks for a validated test. A LAL test follows an established method of a recognized pharmacopoeia, after extraction from the implant by a validated method; the standard names European Pharmacopoeia 2.6.14 and USP <85>. The limit is fixed by the standard: not more than 20.0 Endotoxin Units per implant. The standard notes that orthopaedic implants are not usually in contact with cerebrospinal fluid, and other limits can apply if they are. AAMI ST72 is named for help with methods, levels and the effect of sterilization on endotoxin.


Organic residues: TOC and THC


Clause 5.5.1 says the method is chosen from at least the type of contamination possible, the implant materials and the sensitivity of the method. Two methods carry most of the load:


  • TOC (total organic carbon), or another suitable method, for water-soluble organic contaminants.

  • THC (total hydrocarbons), or another suitable method, for hydrocarbons soluble in nonpolar solvents, unless there is no risk of hydrocarbon contamination.


Tests may use sterilized or non-sterilized specimens; the standard notes sterilization can change the type and amount of organic residue. Some implant materials release organics into the solvent themselves, so procedures may be needed to tell material from contaminant.


Where do the limits come from? Clause 5.5.1 says biological evaluation data from devices with known organic contamination, made with similar materials and processes, may set preliminary or final levels. If a company has no historical TOC or THC data linked to a biological evaluation, the standard notes that the limits in NF S94-091 (0.500 mg per implant for TOC and 0.500 mg per implant for THC) can serve as a starting point.


Two cautions come with that. TOC and THC detect a wide range of organics but do not identify them, so they show that the process is under control, not that a specific residue is safe. And specific organic contaminants can be critical well below the NF S94-091 values. Type and amount both matter. Our guides to chemical characterization under ISO 10993-18 and toxicological risk assessment under ISO 10993-17 cover how specific substances are assessed.


For detection, clause 5.5.3 names EN 1484, European Pharmacopoeia 2.2.44 or USP <643> for TOC, and gas chromatography per ISO 9377-2, FT-IR per ASTM D7066-04 or another validated method for THC.


Inorganic residues


Clause 5.6 says the inorganic contaminants likely to be present are identified in the risk assessment, and those found critical are measured after exhaustive extraction with water, acidified water or another justified solvent. Suitable methods named include ICP with atomic or optical emission spectrometry, ICP-MS, ion chromatography and ion chromatography with MS. If water is the solvent, the extract's conductivity can be a first screen.


The limits come from the biological effects of each inorganic contaminant. The standard points to ISO 10993-17 for allowable limits and ICH Q3D for elemental impurities.


Particulates and cytotoxicity


If cleanliness for particulates is to be shown, clause 5.7 says the type and sensitivity of the test come from the risk management results, and any extraction must avoid making particles through mechanical force. AAMI TIR42 is named for help with methods and levels.


Cytotoxicity is determined by an ISO 10993-5 method on sterilized specimens (clause 5.8). Any cytotoxic effect is investigated for its cause. If it comes from contamination the cleaning did not remove, or from the cleaning itself, control measures are put in place. The standard notes a cytotoxic effect may have other causes.


Exhaustive extraction


A residue left on the implant after extraction is a residue the test never saw. Clause 3.11 defines exhaustive extraction as extraction, usually in several steps, until the amount released in a later step is less than 10 % of the amount released in the first step.


For organic and critical inorganic residues, clauses 5.5.2 and 5.6 require:


  • The right solvent. Water pure enough not to interfere for water-soluble organics. Non-polar solvents for hydrocarbons, chosen so they do not decompose or leach from the material: hexane or halogenated solvents if compatible, and propyl alcohol is also possible for UHMWPE and PEEK. Water, acidified water or another justified solvent for inorganics.

  • Equipment that does not interfere with the solvent.

  • Justified extraction conditions. ISO 10993-12, ASTM F2459 and ASTM G136 are named as sources of methods.

  • Proof of exhaustiveness. If the first extract is already below the quantitation limit, exhaustiveness can be taken as shown.

  • Verification on the worst-case specimen before testing on the cleaning family is finalized.


A made-up example: the first water extraction of a worst-case screw gives 0.21 mg TOC and the second gives 0.015 mg. That is 7 % of the first, below 10 %, so the extraction is exhaustive.


Validating the test methods


Clause 4.7 says every test method used to show conformity is validated and documented. Where applicable, the record covers:


  • why the method suits the contaminants that can be present;

  • extraction efficiency;

  • detection limit, quantitation limit and accuracy;

  • extraction blanks and reference materials; and

  • whether the method can show conformity to the acceptance criteria.


ICH Q2(R1) is named for the validation of analytical procedures, and ISO/IEC 17025 for laboratory competence. Clause 4.2 asks the same question from the risk side: can the methods measure each contaminant at the level needed? A quantitation limit close to the acceptance limit leaves no room to see a trend.


Putting it into practice


The Implant Cleaning Validation Kit includes a Residue Limit Rationale form with a guidance column for each test, preliminary and final criteria, and a Test Method Validation Record covering solvents, extraction conditions, exhaustiveness, blanks, and detection and quantitation limits.


Frequently asked questions


Is 0.500 mg TOC per implant the ISO 19227 limit?


No. ISO 19227 notes that the NF S94-091 values of 0.500 mg per implant for TOC and for THC can serve as a starting point when a company has no historical data linked to a biological evaluation. It also warns that specific organic contaminants can be critical well below them.


What is the endotoxin limit for orthopaedic implants?


Not more than 20.0 EU per implant when endotoxin reduction is a purpose of the cleaning. Other limits can apply to implants in contact with cerebrospinal fluid.


How do I prove an extraction is exhaustive?


Extract the same specimen again until a later step releases less than 10 % of what the first step released, or show the first extract is below the quantitation limit. Do it on the worst-case specimen before family testing is finalized.


Must cytotoxicity specimens be sterilized?


Yes. ISO 19227 says cytotoxicity specimens are sterilized, by the terminal sterilization method or, for implants delivered non-sterile, the method recommended to the user.