EO Half-Cycle Overkill Validation: Fractional Cycles, PCDs and Biological Indicators Under ISO 11135

11 hours ago
6 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


The microbiological performance qualification (MPQ) is the part of an ethylene oxide validation that proves the process kills. Under ISO 11135 there is more than one way to do it, and the choice decides how many runs you need, what a pass looks like and how your routine exposure time is set. This guide explains one of them, the overkill half cycle, and the decisions around it: which approach to use, how to choose a process challenge device (PCD), how many biological indicators and sensors to plan for, and how to read the results.


Each summary of the standard is in our own words and refers to ISO 11135:2014 with Amendment 1:2018. For the whole validation route, see EO Sterilization Validation Under ISO 11135: A Step-by-Step Route.


Three ways to show lethality


ISO 11135 clause 8 lets you determine the rate of microbial inactivation in one of three ways:


  • The BI/bioburden approach (Annex A): exposure time is set from the biological indicator's lethal rate (its D-value) together with the population and relative resistance of the product bioburden, so an SAL can be predicted. It needs bioburden that is relatively consistent over time and no more resistant than the indicator.

  • The overkill approach (Annex B): the process is shown to deliver far more lethality than the bioburden needs, either by the half cycle or by cycle calculation.

  • An alternative method that shows the required SAL is reached.


The overkill approach does not set exposure time from the bioburden population, but it still requires you to show that the bioburden is less resistant than the biological indicator in the internal PCD. It is not a way to avoid knowing your bioburden.


The overkill half cycle in brief


Annex B describes two overkill methods. In the half-cycle method:


  1. Run a short fractional cycle in which some biological indicators survive. Recoverable survivors show that the recovery technique works.

  2. Run three consecutive half cycles that each give total inactivation of biological indicators of at least 10^6, placed in a PCD where appropriate. This confirms the minimum exposure time.

  3. Set the routine exposure time at no less than double that minimum.


The other overkill method is cycle calculation: routine parameters must deliver at least a 12 spore log reduction of the biological indicator, found by direct enumeration or the fraction-negative method.


Clause 9.4.2 adds that if process definition was done in a developmental chamber, the MPQ in the production chamber includes at least three fractional or three half cycles. MPQ parameters are chosen to deliver less lethality than the routine process.


Choosing the process challenge device


A biological indicator on its own tells you little. Clauses 8.6 and 8.7 ask that indicators used to establish the process meet ISO 11138-2 (and ISO 11138-1 if commercially supplied), are at least as resistant to EO as the product bioburden, and are placed in a PCD. The PCD must challenge the process at least as much as the natural bioburden at the most difficult-to-sterilize location in the product.


That gives two kinds of PCD:


  • Internal PCD: product with an indicator placed at, or related to, the hardest location. Annex B asks for the PCD to sit where sterilizing conditions are most difficult to achieve, or for its relationship to that location to be established.

  • External PCD: a device outside the product, often used for routine release. If a different PCD is used for routine release than in the MPQ, it should be at least as resistant as the MPQ PCD (clause 10.3).


For example, the hardest location may be deep inside a lumen, under a closed clamp or between mated surfaces. Write down why you chose it. When product changes later, clause 12.4 asks you to reconfirm, as needed, that the internal and external PCDs are still right against the product bioburden.


How many indicators and sensors


ISO 11135 Annex C gives informative guidance, not requirements, on numbers. Rounding up in each case:


  • MPQ biological indicators: three per cubic metre of load up to 10 cubic metres, with a minimum of five, plus one per extra cubic metre above 10.

  • Routine biological indicators, if used: half the MPQ number, up to 30.

  • PQ temperature sensors: at least one per cubic metre of product load, minimum three.

  • PQ humidity sensors: in the packaging within the load, about one per 2.5 cubic metres, minimum two.


For a load of 1.6 cubic metres, for example, that works out to five internal PCDs (1.6 x 3 = 4.8, so the minimum of five), three temperature sensors and two humidity sensors. Your sterilizer may plan more. Record the numbers and the reasoning in the master plan.


Reading the results


What counts as a pass depends on the approach (clause 9.4.2):


  • Overkill half cycle: no internal PCD may be positive. A positive external PCD is acceptable if the external PCD has been shown to be more resistant than the internal ones.

  • Cycle calculation or BI/bioburden approach: some internal PCDs may survive, but the calculated SAL must meet the specified value.


Recovery conditions matter as much as the cycle. Annexes A and B both ask that indicator recovery and incubation conditions are established and documented, allowing for the delayed outgrowth of spores exposed to EO. Agree the incubation conditions with the laboratory before the first run.


If a test of sterility is performed on product, for example in the fractional cycle to compare product with the PCD, it follows ISO 11737-2 (clause 8.9).


A worked example


The numbers below are made up to show how the decisions connect.


A single-use tubing set has a 3 m lumen with a closed clamp. No validated family has a lumen that long, so the product is a greater challenge and needs full validation. The hardest location is the middle of the lumen at the clamp. The internal PCD is an indicator strip in the lumen at that point. The external PCD is an indicator in a tube outside the shipper.


The load is 1.6 cubic metres, so the plan calls for five internal PCDs per MPQ run. In the fractional cycle, three of five internal PCDs are positive and none of ten product sterility samples are positive. The recovery technique works and the PCD is more resistant than the product bioburden. In three consecutive half cycles at 30 minutes of exposure, none of the five internal PCDs are positive. The routine exposure time is set at 60 minutes.


After the MPQ


The MPQ sets lethality. It does not show reproducibility. That is the job of the physical PQ: at least three consecutive full routine cycles that meet the acceptance criteria throughout the load (clause 9.4.3). If the PPQ runs alongside at least three MPQ runs, at least one more run uses the full routine process.


Saleable product used in an MPQ needs a full-exposure sterilization process and a formal review of acceptance before it can be released to market (clause 11.4).


How the A.P. Lyon kit fits


The EO Sterilization Validation Kit includes a Product Family and Process Challenge Device Record for the hardest location and PCD choice, a validation master plan with the Annex C numbers worked out, and a Microbiological PQ Protocol and Report covering the fractional cycle, three half cycles and the exposure time decision. Every step names its owner and record, with the clause reference in a separate column.


Frequently asked questions


Why is the routine exposure time double the half cycle?


Annex B's half-cycle method sets the routine exposure time at no less than twice the minimum exposure time confirmed by three half cycles with total inactivation.


Is a positive external PCD in a half cycle a failure?


Not necessarily. It is acceptable if the external PCD has been shown to be more resistant than the internal PCDs. A positive internal PCD in a half cycle is a failure.


Are the Annex C sensor and BI numbers mandatory?


No. Annex C is informative guidance. Record the numbers you use and why.


Do I still need bioburden data with the overkill approach?


Yes. The bioburden must be shown to be less resistant than the indicator in the internal PCD.