Dose Mapping and Maximum Acceptable Dose Under ISO 11137: Fitting the Dose Window to Your Load

11 hours ago
6 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.


Every radiation sterilization process lives inside a window. At the bottom is the sterilization dose: no part of the product may get less. At the top is the maximum acceptable dose: no part of the product may get more. Between them sits the spread of dose the irradiator actually delivers through a real load. If that spread is wider than the window, the process cannot work as loaded, however good the microbiology is.


This guide covers the top of the window and the spread: how to establish the maximum acceptable dose, how product dose mapping works, and how the two come together in a routine monitoring dose range. Each summary is in our own words and refers to ISO 11137-1:2025 and the guidance in ISO 11137-3:2017. For the whole route, see Radiation Sterilization Validation Under ISO 11137-1:2025. For the bottom of the window, see VDmax25 Dose Substantiation and Dose Audits.


Part 1: The maximum acceptable dose


ISO 11137-1 clause 8.1 asks you to establish the maximum acceptable dose so that product treated at that dose still meets its functional requirements throughout its specified lifetime. Three things are needed:


  • A facility able to assess product and packaging against acceptance criteria.

  • Product in its sterile barrier system, representative of routine production.

  • An irradiator that delivers precise and accurate doses.


In practice the study runs in four steps:


  1. Design the study. Choose doses at and above the highest dose you expect in routine processing. Choose the product and package tests and their acceptance criteria.

  2. Irradiate. The irradiator treats the samples and records the doses actually delivered.

  3. Test, fresh and aged. Test product function and package integrity right after irradiation and again after real-time or accelerated aging to the end of the labeled shelf life. The maximum acceptable dose must hold throughout the product's lifetime, so the aged results matter.

  4. Set the maximum acceptable dose at the highest dose where every test passes.


If the maximum acceptable dose will be used on a different radiation source from the one used in the study, clause 8.4.1 asks for a documented assessment that differences in irradiation conditions do not affect its validity.


Part 2: What the irradiator has already mapped


Before mapping your product, check the irradiator's operational qualification (clause 9.2). OQ dose mapping uses containers filled to their upper limit with homogeneous material, with dosimeters at known depths. For gamma and X-ray, there are enough surrounding containers to mimic a full irradiator. Mapping covers each conveyor path and container type, on enough containers to find variability, with at least three replicates recommended. OQ also determines the effect of a process interruption on dose and relates the process parameters to dose.


ISO 11137-3 guidance adds the dosimetry checks: dosimeters calibrated in absorbed dose to water under conditions that match actual use, preferably in the irradiator of intended use; traceability through a recognized national metrology institute or an ISO/IEC 17025 laboratory; and verification after significant changes such as source replenishment. A supplier's calibration curve is not used without verification.


Part 3: Mapping your product


Performance qualification dose mapping (ISO 11137-1 clause 9.3) uses product, or simulated product, in a specified loading pattern. Start by specifying the presentation:


  • Package dimensions and density.

  • Orientation of the product in the package, and of the package in the container.

  • Container type and any positioning or attenuating materials.

  • The conveyor path.


Then map the dose for each processing category, each conveyor path and each container type, on enough containers to find variability, again with at least three replicates recommended. Determine the effect of partially filled containers. For gamma and X-ray, also determine the effect of different densities being present together.


Part 4: Reading the map


The map has three jobs:


  1. Find the zones and values of the minimum and maximum dose in the load.

  2. Choose a routine monitoring position where a dosimeter can be placed on every run, and relate the minimum and maximum doses to the dose there.

  3. Relate the process parameters to the routine monitoring dose.


The ratio of maximum to minimum dose shows how wide the spread is. It is the number you compare with the gap between the sterilization dose and the maximum acceptable dose. If the spread does not fit, ISO 11137-1 clause 8.3 leaves the fixes to you: change the loading pattern, the packaging or the irradiator, or redesign the product.


Part 5: Setting the routine monitoring dose range


Now set the acceptable range for the dose at the routine monitoring position. Using the relationships from the map:


  • The lower limit is set so the minimum dose in the load is at least the sterilization dose.

  • The upper limit is set so the maximum dose in the load is no more than the maximum acceptable dose.


Take dose measurement uncertainty into account here. ISO 11137-3 guidance says measured doses are the best estimate and are not corrected by their uncertainty; instead, uncertainty is considered when setting the acceptable range at the routine monitoring position.


A worked example


The numbers below are made up to show how the parts connect.


A family of polypropylene syringes is gamma sterilized. In the maximum dose study, samples at 45 kGy pass plunger force, leak and pouch seal tests after accelerated aging to the 3-year shelf life; samples at 55 kGy show brittleness. The maximum acceptable dose is set at 45 kGy. VDmax25 substantiates 25 kGy.


Three dose mapping runs with shippers in the routine loading pattern put the minimum dose at the centre of the shipper and the maximum at the outer face, with a maximum to minimum ratio of 1.4. A load that receives 25 kGy at its minimum therefore receives about 35 kGy at its maximum, well inside the 45 kGy limit. The routine monitoring position is on the outside of the carrier, and its range is set from the map so the minimum stays at or above 25 kGy and the maximum at or below 45 kGy, allowing for measurement uncertainty.


Part 6: Keeping the map valid


ISO 11137-3 guidance says PQ is repeated when OQ shows the irradiator has changed, or when the product changes in a way that may affect dose distribution. ISO 11137-1 clause 12.5 asks for any irradiator change that could affect dose or its distribution, such as source replenishment or a conveyor change, to be assessed, with IQ, OQ or PQ repeated as needed. Product, package and presentation changes are assessed against the process specification before they take effect.


How the A.P. Lyon kit fits


The Radiation Sterilization Validation Kit includes a Maximum Acceptable Dose Study form, a Dose Mapping Protocol and Report that reviews the irradiator's IQ, OQ and dosimetry evidence and records product mapping and the routine dose range, and a Validation Report and Release Specification. Every step names its owner and record, with the clause reference in a separate column.


Frequently asked questions


Who does dose mapping, the manufacturer or the irradiator?


In the A.P. Lyon kit, the irradiator runs it and the manufacturer specifies the product presentation and approves the result. Write the split into your agreement.


Can I correct measured doses for their uncertainty?


No. ISO 11137-3 guidance says measured doses are not corrected by their uncertainty. Allow for uncertainty when setting the routine monitoring dose range.


How many dose mapping runs are needed?


ISO 11137-1 recommends at least three replicates for each processing category, conveyor path and container type.


Why test aged samples for the maximum acceptable dose?


Because the maximum acceptable dose must keep the product within its functional requirements throughout its specified lifetime, not just on the day of irradiation.