VDmax25 Dose Substantiation and Dose Audits Under ISO 11137-2: Samples, Results and Failed Audits
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.
If you sterilize at 25 kGy, the evidence that 25 kGy is enough can come from a small experiment called VDmax25. The same experiment, repeated as a dose audit, then shows every quarter that the dose still works. This guide walks through both: how to run VDmax25, how to read it, how dose audits work, when the audit interval can be extended, and what to do when an audit fails.
Each summary of the standards is in our own words and refers to ISO 11137-2:2013 with Amendment 1:2022 and ISO 11137-1:2025. For the whole validation route, see Radiation Sterilization Validation Under ISO 11137-1:2025.
Setting a dose versus substantiating one
ISO 11137-2 clauses 6.1 to 6.3 give two routes:
Set a product-specific dose from bioburden, by Method 1, Method 2A or 2B, or an equivalent method.
Select a dose and substantiate it. For 25 kGy, use VDmax25 (average bioburden up to 1,000), or Method 1 or 2 giving 25 kGy or less at an SAL of 10^-6. For 15 kGy, use VDmax15 (average bioburden up to 1.5).
Amendment 1 adds other selected doses. A dose of 17.5, 20, 22.5, 27.5, 30, 32.5 or 35 kGy can be substantiated by Method VDmaxSD of ISO 13004, by Method 1 or 2, or by an equivalent method (clause 6.4).
VDmax25 is the route this guide follows. It uses few test items (a VDmax dose audit needs 20 items, against 110 for a Method 1 audit), but it only answers one question: is 25 kGy enough?
Step 1: Choose the samples
Clauses 5.2 to 5.5 set the rules for samples:
Product represents routine production, each item preferably from a separate package.
The delay between sampling and the verification dose experiment reflects the normal time from the last manufacturing step to sterilization. Rejects may be used if fully processed.
If you test a sample item portion (SIP) instead of the whole product, show that it is adequate.
Bioburden is measured on packaged product. Do not manipulate product in ways that change its bioburden or its response to radiation. Product may be taken apart and repacked in radiation-tolerant materials.
Bioburden and sterility tests follow ISO 11737-1 and -2.
Step 2: Determine the bioburden
Clauses 9.2.1 to 9.2.4 ask for bioburden on 10 items from each of three batches, with the correction factor applied. Then decide which average to use:
If any batch average is two or more times the overall average, use the highest batch average.
Otherwise, use the overall average.
VDmax25 can be used only if every batch average is 1,000 or less. The whole product must be tested if the average is 0.9 or less; an SIP is allowed above 0.9. Read the verification dose for your average from the VDmax25 table in ISO 11137-2.
Step 3: Run the verification dose experiment
Irradiate 10 items from one batch at the verification dose (clause 9.2.5). The irradiator must be qualified under ISO 11137-1, with enough dose mapping to find the highest and lowest doses to the product. Two dose checks apply:
The highest dose must not be more than 10 % above the verification dose. If it is, repeat.
The mean of the highest and lowest doses should be at least 90 % of the verification dose.
Then test each item for sterility under ISO 11737-2.
Step 4: Read the result
Clause 9.2.6 gives a simple rule:
0 or 1 positive: 25 kGy is substantiated.
2 positives: run a confirmatory experiment.
3 or more positives: 25 kGy is not substantiated. If the positives come from a laboratory error, a dose delivery error or a specific bioburden cause, correct it and repeat with 10 new items. Otherwise choose another method.
Record the result with the bioburden data, the doses delivered and the sterility test records. Then specify the sterilization dose alongside the maximum acceptable dose (ISO 11137-1 clause 8.3).
Dose audits: how often
A dose audit repeats the verification dose experiment to show the sterilization dose is still appropriate. ISO 11137-1 clause 12.1.3 asks for audits at least four times a year with no more than four months between them, or at an interval set by a documented rationale. The results apply to every product in the family. No audit is needed when product is not being made; if batches are made less often than the audit interval, audit the next batch.
Dose audits: how to run them
ISO 11137-2 clauses 10.1 to 10.3 describe the audit:
For a VDmax dose: take 20 items from one batch. Determine bioburden on 10. Irradiate the other 10 at the original verification dose, with the highest dose no more than 10 % above it (or 0.1 kGy for VDmax15). Test each for sterility in the original media and conditions. 0 or 1 positive accepts the audit; 2 positives call for a confirmatory audit.
For a Method 1, 2A or 2B dose: take 110 items, determine bioburden on 10, irradiate 100 at the verification dose and test each. Up to 2 positives accepts.
Use the same SIP as the original exercise. The audit bioburden data are for trending and investigations, not for setting a new verification dose.
Extending the audit interval
ISO 11137-1 clauses 12.1.3.2 and 12.1.3.3 allow the interval to be lengthened to no more than 12 months, but only when all of these hold:
At least four consecutive audits at the current interval needed neither dose augmentation nor re-establishment.
Bioburden data and characterization show stability within the action level.
Bioburden control is demonstrated.
The rationale must review and record at least: alert and action levels; bioburden data, period and characterization; resistance data; the dose setting method; the margin between routine dose and sterilization dose; materials, especially those of natural origin; manufacturing steps that affect bioburden; process, environment and personnel controls; time between batches; and data on similar products. Quarterly bioburden testing continues either way.
When an audit fails
First ask whether the positives come from a sterility test error, a dose delivery error or a specific bioburden cause. If so, correct it and repeat the verification dose experiment at the same dose as soon as practicable (ISO 11137-2 clauses 10.2.5.2, 10.2.5.3 and 10.3.5.3).
If not, the action depends on the method and the number of positives. For a VDmax dose:
3 to 6 positives: augment the dose immediately and re-establish it by another method, continuing at the augmented dose meanwhile.
7 or more positives: stop sterilizing at that dose, with no augmentation, until it is re-established by another method.
For Method 1 or 2 doses the thresholds differ: 3 to 4 positives, 5 to 15, and more than 15.
Whatever the outcome, investigate the cause, take correction or corrective action, risk-assess batches processed since the last good audit for suitability for use, and record it all (clause 10.4). Audits return to at least four a year until the cause is corrected and the extension criteria are met again (ISO 11137-1 clause 12.1.3.4).
When bioburden goes high
A bioburden result above the action level is investigated and the organisms characterized (ISO 11137-1 clause 12.1.2.4). A sampling or laboratory error allows a retest. If the high result is confirmed and differs from routine, run a dose audit on the same or next batch. If the audit passes but the average stays above the action level, processing may continue at the current dose, with audits at least four times a year until bioburden falls back or the dose is re-established.
How the A.P. Lyon kit fits
The Radiation Sterilization Validation Kit includes a Dose Setting Protocol and Record for VDmax25 and Method 1, a Sterility Test Record, and a Dose Audit Record and Change Review with the interval rationale checklist, plus the procedure steps for each decision above. Every step names its owner and record, with the clause reference in a separate column.
Frequently asked questions
What does VDmax25 prove?
That a selected dose of 25 kGy is appropriate for product with your bioburden. It does not set a lower product-specific dose; Method 1 or 2 does that.
What bioburden is too high for VDmax25?
Any batch average above 1,000. Then use another method.
Does a passing dose audit need any follow-up?
Record it, trend the audit bioburden, and keep the schedule. Four consecutive passes without augmentation are one of the conditions for extending the interval.
Does an audit result apply only to the product tested?
No. Dose audit results and actions apply to every product in the family.