Adverse Event and Device Deficiency Reporting in Device Clinical Investigations: ISO 14155 and 21 CFR 812
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 28, 2026.
Safety reporting is where a clinical investigation is most often found wanting. An event is recorded in the notes but never reaches the sponsor. A device that failed is thrown away instead of returned. The investigator and the sponsor disagree about whether an event was device related, and nobody tells the ethics committee. A report goes in on day 14 when the CIP said 24 hours. Each of these is avoidable with a clear route and a log that shows what is due.
This guide sets out that route under ISO 14155:2026 and, for U.S. investigations, 21 CFR Part 812. The requirements are summarized in our own words; check the standard and the official text at www.ecfr.gov before relying on them.
The terms that drive every decision
ISO 14155:2026 defines the categories that decide what is reported, by whom and how fast.
Adverse event (AE): any untoward medical occurrence, unintended disease or injury, or untoward clinical sign (including abnormal laboratory findings) in subjects, users or other persons, whether or not related to the investigational device.
Adverse device effect (ADE): an adverse event related to the use of the investigational device, including events caused by inadequate instructions for use, malfunction, use error or intentional misuse.
Serious adverse event (SAE): an adverse event that led to death; to serious deterioration in health (a life-threatening illness or injury, permanent impairment of a body structure or function, in-patient or prolonged hospitalization, or medical or surgical intervention to prevent these); or to foetal distress, foetal death, congenital abnormality or birth defect. Planned hospitalization for a pre-existing condition, or for a procedure the CIP requires, without serious deterioration in health, is not an SAE.
Serious adverse device effect (SADE): an adverse device effect with any of the consequences of a serious adverse event.
Unanticipated serious adverse device effect (USADE): a SADE whose nature, incidence, severity or outcome has not been identified in the current risk assessment. An anticipated one (ASADE) has been.
Device deficiency: an inadequacy in the identity, quality, durability, reliability, usability, safety or performance of a device, including malfunctions, use errors and inadequate labelling.
Serious health threat: a signal from any adverse event or device deficiency that indicates an imminent risk of death or serious deterioration in health and needs prompt remedial action for others.
Annex F of ISO 14155 (informative) shows how the categories fit together and gives categorization charts for adverse events and device deficiencies.
Step 1: record every event and every deficiency
ISO 14155 clause 7.4.2 requires all adverse events, and any new information about them, to be documented in a timely manner throughout the investigation. Clause 10.8 a) requires the principal investigator to record every adverse event and every observed device deficiency, with an assessment.
"Every" is the important word. A sore throat in a wound care study is still an adverse event. Recording it costs a minute; not recording it means the safety data in the final report is incomplete. ISO 14155 allows reduced recording or reporting only where the CIP identifies and justifies it and the ethics committee (and, as required, the regulatory authority) approved it in advance.
In the United States, 21 CFR 812.140(a)(3) requires each investigator to keep a case history for each subject, including records of adverse device effects.
Step 2: check seriousness against the criteria
For each event, go through the SAE criteria one by one and record the answer. Checking every criterion, rather than writing "serious: yes", makes the decision checkable later, and it catches the common error of treating a planned admission as an SAE or missing that a prolonged stay made an event serious.
Step 3: assess the relationship
The investigator assesses whether the event is related to the investigational device, to a procedure the CIP requires that is outside routine clinical practice, and to the comparator where there is one. ISO 14155 keeps device risks and procedure risks apart (clauses 6.2.1 and 7.4.5), so the relationship to each is assessed separately. Then choose the category, using Annex F as guidance.
Step 4: handle device deficiencies as their own events
ISO 14155 clause 7.4.3 requires every device deficiency of the investigational device to be documented, whether or not it caused harm, and managed by the sponsor under its procedures for nonconforming product, with corrective and preventive action where needed. Deficiencies of a comparator are documented too.
Two things matter most at the site:
Decide whether it could have led to a SADE. The test has three parts: if suitable action had not been taken, if intervention had not been made, or if circumstances had been less fortunate. A deficiency that meets the test is reported like a serious event, even though nobody was hurt.
Keep the device. The sponsor arranges its safe return, and analysis of used or explanted devices supports the report where applicable. A device discarded at the site cannot be analyzed.
Step 5: report to the sponsor on time
Clause 10.8 b) requires the principal investigator to report to the sponsor, without unjustified delay, all serious adverse events and all device deficiencies that could have led to a SADE, followed promptly by detailed written reports as the CIP specifies. Clause 10.8 c) and d) add reports to the ethics committee if the CIP or the committee requires it, and to regulatory authorities where required.
ISO 14155 does not set a number of hours or days. The CIP, the ethics committee and national regulations do. Write the timelines into the CIP and into the site's working tools, so the site coordinator can see the due date the moment an event is entered.
In the United States, 21 CFR 812.150(a)(1) sets a fixed limit: an investigator reports an unanticipated adverse device effect to the sponsor and the reviewing IRB within 10 working days of first learning of it.
If an event may be a serious health threat, the site tells the sponsor at once. ISO 14155 clause 7.4.1 says such signals can be detected by either party but are evaluated by the sponsor.
Step 6: the sponsor reviews and decides
ISO 14155 clause 9.2.5 makes the sponsor responsible for classifying adverse events and for ongoing safety evaluation. The sponsor:
reviews the investigator's assessment of every adverse event and records in writing its own decision on seriousness and relationship;
reviews every device deficiency and records whether it could have led to a SADE;
where it disagrees with the investigator, communicates both opinions to the ethics committee, regulatory authorities and data monitoring committee as required;
reports serious adverse events and such device deficiencies to the ethics committee and regulatory authorities as required;
in a multicentre investigation, informs all principal investigators in writing of serious adverse events at all sites, within a time frame based on the risk assessment;
tells ethics committees and regulatory authorities about significant new information;
for SADEs and deficiencies that could have led to one, decides whether the risk analysis needs updating and whether corrective or preventive action is needed.
A clinical events committee can do the classification independently to reduce bias.
In the United States, 21 CFR 812.46(b) requires the sponsor to evaluate any unanticipated adverse device effect immediately. Under 21 CFR 812.150(b)(1), the sponsor reports the results of that evaluation to FDA, all reviewing IRBs and all participating investigators within 10 working days of first receiving notice of the effect. If the effect presents an unreasonable risk to subjects, the sponsor terminates all or the affected parts of the investigation within 5 working days of making that determination and no later than 15 working days after first receiving notice.
Step 7: keep one log and watch the dates
Each report on its own is not enough. The sponsor and the monitor need a log of every event and deficiency that shows:
the category given by the investigator and by the sponsor;
the date each report was due and the date it was sent;
which reports are late;
which items are still open.
The same log gives the counts for monitoring reports, progress reports to ethics committees and the safety section of the clinical investigation report. ISO 14155 clause 7.4.2 requires all adverse events per the CIP to be reported in an interim or final report of the investigation.
Step 8: close the loop into the quality system
A device deficiency in a clinical investigation is a nonconformity of the investigational device. Raise it under the nonconforming material procedure. Where the deficiency or a SADE points to a cause in the design or manufacture, open a CAPA. Update the risk analysis where the evaluation shows it is needed.
If an unanticipated effect or a suspected serious health threat makes the risk unacceptable, ISO 14155 clause 8.2.2 requires the sponsor to suspend the investigation while the risk is assessed, in proportion to the risk. The sponsor terminates it if an unacceptable risk that cannot be controlled is confirmed.
Common mistakes
Only serious events recorded. Every adverse event is recorded, serious or not.
Device deficiencies recorded only when someone was hurt. The "could have led to a SADE" test exists for near misses.
The device is discarded. It must go back to the sponsor for analysis.
Timelines live only in the CIP. The site cannot see what is due until it is late.
Sponsor disagreement kept internal. Both opinions go to the concerned parties.
No link to CAPA. The same deficiency appears at three sites before anyone looks for a cause.
Putting it into practice
The Clinical Investigation Conduct Forms Kit turns this route into numbered steps. FORM 0104-CIF records one adverse event or device deficiency with the seriousness criteria, relationship, the "could have led to a SADE" test, the category, the reporting dates and the sponsor's review. FORM 0105-CIF is an Excel log that calculates the due date at the sponsor from your CIP timelines and the 10-working-day U.S. date for an unanticipated adverse device effect, flags late reports and counts events by category. It comes in ISO 13485 and dual-standard editions and builds on the A.P. Lyon Clinical Investigation Procedure (SOP 7-039 or SOP 7-039-C). For how monitoring catches missed reports, see How to Write a Risk-Based Monitoring Plan Under ISO 14155:2026. For CAPA on a device deficiency, see Nonconforming Material, Failure Investigation and CAPA.
Frequently asked questions
Does ISO 14155 set reporting timelines?
No. ISO 14155:2026 requires serious events and device deficiencies that could have led to a SADE to be reported to the sponsor without unjustified delay, and leaves specific timelines to the CIP, the ethics committee and national regulations. In the United States, 21 CFR 812.150 sets fixed limits, such as 10 working days for an unanticipated adverse device effect.
Is a device deficiency with no harm reportable?
It is always recorded. It is reported like a serious event if it could have led to a serious adverse device effect had suitable action not been taken, had intervention not been made, or had circumstances been less fortunate.
Who decides whether an event is device related?
The investigator assesses it first. The sponsor then reviews that assessment and records its own decision in writing. If they disagree, both opinions are communicated to the ethics committee, regulatory authorities and data monitoring committee as required.
Does ISO 14155 require a coding dictionary such as MedDRA?
No. ISO 14155:2026 says the sponsor should consider using standardized terminology to code adverse events and device deficiencies, but it does not name a dictionary.