Design Inputs and the Trace Matrix: Turning User Needs Into Requirements You Can Verify
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.
Weak design inputs cause trouble at every later stage. A requirement that cannot be measured cannot be verified. A user need with no requirement behind it is never designed for. A risk control that never became a requirement is never tested. Most of these problems show up late, at the verification review or the final review, when they are expensive to fix.
This guide covers the front end of design control: user needs, design inputs, the requirement check and the trace matrix that holds them together. Each summary of a requirement is in our own words. For the whole route, see Design Controls Step by Step Under ISO 13485 and the FDA QMSR.
What ISO 13485 asks for
Clause 7.3.3 asks that inputs relating to product requirements are set and recorded, and that they cover:
functional, performance, usability and safety requirements, according to the intended use
the regulatory requirements and standards that apply
the outputs of risk management that apply
information from earlier similar designs, where useful
other requirements essential to the design of the product and its processes
The inputs are reviewed for adequacy and approved. Each requirement must be complete, unambiguous, able to be verified or validated, and not in conflict with another. The clause points to IEC 62366-1 for usability.
Clause 7.3.2 adds that the design plan records the methods used to trace design outputs back to design inputs. That is the job of the trace matrix.
Start with user needs
A user need is what the user, the patient or another stakeholder needs from the device, in their words, for the intended use. "The clinician can start and stop suction with one thumb" is a user need. It cannot be tested as written, and it should not be. User needs are confirmed by design validation, with real or simulated users.
Write the intended use, the users, the patients and the use environment first. Then list the needs, give each an ID such as UN-01, and record where it came from: user research, complaints on an earlier device, clinical input or marketing.
Turn each need into design inputs
A design input is a physical or performance requirement of the device that the design is built on. Each user need becomes one or more inputs, each with an ID such as DI-07, a parent user need, acceptance criteria and the planned verification method.
The thumb-control need above might become "the vent closes fully under a thumb force of 2 N or less", verified by test at the conditions of use. The number is the point: it gives the engineer a target and the tester a pass line.
Not every input comes from a user need. Regulatory requirements and standards come from Regulatory Affairs. Risk controls come from the risk analysis: under ISO 14971, risk control options are used in priority order, starting with inherently safe design, so many controls become design inputs. Record the hazard ID against each one. Packaging, shelf life, sterility, labeling, manufacturing and service needs also become inputs.
Check every requirement
Before the input review, check each requirement against the four tests in clause 7.3.3:
Complete: it says everything needed, including conditions and units.
Unambiguous: two engineers would read it the same way.
Able to be verified or validated: it has acceptance criteria and a method.
Not in conflict with another requirement.
Conflicts are normal. A soft tip and enough stiffness to advance a catheter pull against each other. The check is where they are found and settled, with the resolution recorded, rather than at the verification bench.
Build the trace matrix
The trace matrix is one row per user need and one row per design input. For each row it records:
the ID, the type (user need or design input) and, for an input, its parent user need
the source: user need, regulatory, standard, risk control, previous design or other
the requirement and its acceptance criteria
whether it is an essential characteristic for safe and proper use (clause 7.3.4 asks outputs to state these)
the risk control hazard ID, if any
the design output that meets it, with document number and revision
the verification or validation method, the record number and the result
The matrix earns its keep by showing gaps. A user need with no design input, a design input with no output, an input with no verification record, a user need with no validation record, a failed test and an open item each need action. If the matrix calculates a status for every row and counts them, the counts can go in front of every design review.
Use the matrix at every review
Clause 7.3.5 asks reviews to judge whether the results can meet requirements and to propose actions. A one-page count of closed rows, open rows and each kind of gap gives the review that judgement in a form nobody can argue with. By the final review, every count except "closed" should be zero.
When a design input changes after approval, the change goes through design change control under clause 7.3.9, and the matrix row changes with it, so the matrix stays a true picture of the design.
Putting it into practice
The Design Control Kit includes the Design Input Requirements form, with the four-part requirement check, and the Design Trace Matrix as an Excel workbook. The workbook flags each kind of gap on its own and has a summary sheet that says whether the design is ready for the final review. Both are used under the Design Planning and Inputs Procedure, written as numbered steps with an owner for each.
Frequently asked questions
Is a trace matrix required by ISO 13485?
The standard does not name a trace matrix. It asks the design plan to record the methods used to trace outputs to inputs. A trace matrix is the usual way to do that.
Should user needs be verified or validated?
Validated. Design inputs are verified against their acceptance criteria; user needs are confirmed by validation on representative product.
How detailed should a design input be?
Detailed enough that a test can pass or fail it. If two people would judge the same result differently, the requirement is not yet unambiguous.