When Does a Device Need a New UDI Device Identifier? US and EU Rules, GUDID Updates and UDI Records
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 27, 2026.
Assigning the first UDI is the easy part. The harder questions come later: a box changes from 10 units to 5, a brand name changes, a device is relabeled for a distributor. Does it need a new device identifier? Which database has to be updated, and by when? And what records must be kept?
This guide sets out the US rules in 21 CFR Part 830 and the EU rules in Annex VI of Regulation (EU) 2017/745, and a simple way to apply them. Every summary of a requirement is in our own words. For FDA inspection findings on UDI, see FDA 483 Citations for Unique Device Identification.
The two parts of a UDI
A UDI has two parts:
the device identifier (DI, called the UDI-DI in the EU), which identifies the labeler and the specific version or model at a given packaging level
the production identifier (PI, called the UDI-PI in the EU), which carries the lot or batch, serial number, manufacturing date, expiration date or software version, as shown on the label
In the US, 801.40 requires the UDI in plain text and in a machine-readable (AIDC) form, with a PI whenever the label shows a lot, serial number, manufacturing date or expiration date. A class I device UDI need not include a PI (801.30(d)). In the EU, Annex VI Part C 3.5 puts the lot, serial number, software identification or expiry date in the UDI-PI when they appear on the label.
One DI, one version or model
Part 830.40 sets the basic rules. One DI from a given issuing system identifies one version or model, and each DI identifies only one version or model. A discontinued device's DI is never given to another device. It may be reused only if the same device comes back without a change that needs a new DI.
In the EU, the UDI-DI is unique at each level of device packaging (Annex VI Part C 3.4).
When the US requires a new DI
Part 830 names three triggers:
A new version or model (830.50(a)): a change to a device that results in a new version or model.
A new device package (830.50(b)): creating a new package, for example a box of 5 where there was a box of 10.
Relabeling (830.60): a firm that relabels a device that must bear a UDI assigns a new DI and keeps a record linking the prior DI to the new one.
When the EU requires a new UDI-DI
Annex VI Part C 3.9 is broader. A new UDI-DI is needed for any change that could cause misidentification of the device or ambiguity in its traceability, and in particular for any change to:
the name or trade name
the device version or model
the single-use labeling
packaged sterile
the need for sterilization before use
the quantity of devices in a package
critical warnings or contra-indications, for example containing latex or DEHP
Firms that repackage or relabel with their own label keep a record of the original manufacturer's UDI (3.10).
Updating GUDID and EUDAMED
A new DI or UDI-DI needs a new database record. Changes to existing records have their own deadlines:
GUDID: data for a new DI are submitted no later than the date the label must bear the UDI (830.330(a)). Changes are submitted no later than the date a device is first labeled with the changed information, or within 10 business days of the change if the information does not appear on the label (830.330(b)). If FDA questions a record, the labeler corrects it or explains it within 30 days (830.350).
EUDAMED: data for a new UDI-DI are available when the device is placed on the market (Annex VI Part C 5.7). Changes to elements that do not need a new UDI-DI are updated within 30 days (5.8).
Part 830.310 lists what GUDID holds for each DI, including the brand name and version or model as on the label, whether it is labeled sterile or as containing natural rubber latex, MRI safety status, size, the types of PI on the label, the premarket submission number, the listing number, the GMDN term and the number of devices in the package. Each of these should be checked against the approved artwork before it is submitted, since a mismatch between the label and the database is easy to make and easy for an inspector to find.
Keep the UDI records
Part 830.360 requires records of every UDI used and the version or model each DI belongs to, kept for 3 years after the version or model is no longer marketed. Part 830.320(a) requires a named point of contact for device identification. The QMSR adds that the UDI is recorded for each device or batch in its production records (820.35(c)).
A DI register with one row per DI, its packaging level and quantity, its status and dates, and the DI it replaced with the reason, covers the 830.360 and 830.60 records in one place.
A simple decision routine
For every device, package or labeling change:
Ask the US questions: new version or model, new package, or relabeling?
Ask the EU questions: does it change any of the Annex VI Part C 3.9 elements, or could it cause misidentification?
If either answer is yes, assign the new DI or UDI-DI, link it to the old one in the register, and create the database records before the first changed label is used.
If both answers are no, update any changed data elements within the GUDID and EUDAMED deadlines.
Decide whether product already released needs relabeling or correction.
Putting it into practice
The Labeling and UDI Kit includes a UDI Device Identifier Assignment Record, a UDI Database Data Entry Checklist with the 830.310 data elements and the update deadlines, and a Labeling Change Assessment that asks the US and EU new-DI questions for every change. They are used under a step-by-step UDI Assignment and Database Entry Procedure.
Frequently asked questions
Does a new box size always need a new DI?
In the US a new device package needs a new DI under 830.50(b). In the EU a change to the quantity per package needs a new UDI-DI under Annex VI Part C 3.9.
Is a UPC enough for a class I device in the US?
801.40(d) treats a class I device that bears a UPC on its label and packages as meeting the UDI labeling rules, with the UPC serving as its UDI.
Does every device need a UDI?
No. 801.30 lists exceptions, for example research-only, custom, investigational and export devices. A UDI is never required on a shipping container. Record which exception you rely on.