Decision Rules and Statements of Conformity Under ISO/IEC 17025

2 days ago
6 min read

Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 24, 2026.


A customer sends a part and a drawing and asks a simple question: does it pass? When the measured value sits well inside the tolerance, the answer is easy. When it sits close to a limit, the answer depends on measurement uncertainty, and on a rule the lab and the customer should have agreed before any work started.


ISO/IEC 17025:2017 calls that rule a decision rule. This guide explains when the standard requires one, what it has to cover, and how to report a statement of conformity. Each requirement is summarized in our own words.


What a statement of conformity is


A statement of conformity is any statement that an item meets, or does not meet, a specification: pass or fail, in tolerance or out of tolerance, complies or does not comply. Test reports may include one where relevant (clause 7.8.3), and so may calibration certificates (clause 7.8.4).


A decision rule describes how measurement uncertainty is taken into account when the lab makes that statement.


Why uncertainty matters near the limit


Every measurement has uncertainty. Suppose a specification limit is 10.0 mm maximum, the lab measures 9.95 mm, and the expanded uncertainty is ±0.10 mm. The true value could be anywhere from 9.85 mm to 10.05 mm. Calling the part a pass carries a real chance that it is actually out of specification. That chance is the risk of a false accept. The opposite case, calling a good part a fail, is a false reject.


A decision rule decides who carries that risk. The standard does not prescribe a rule; common choices include:


  • Simple acceptance. The measured value is compared directly with the limit. A value inside the limit passes. Uncertainty is reported but does not move the limit.

  • Guarded acceptance. An acceptance limit is set inside the specification limit by a guard band, often based on the expanded uncertainty. A result must fall inside the guard band to pass, which reduces the risk of a false accept.

  • Guarded rejection. The guard band is set outside the specification limit, which reduces the risk of a false reject.

  • A rule set elsewhere. The customer, a regulation or the specification itself sets the rule, and the lab applies it.


The right rule depends on the consequences of a wrong decision. For a safety-critical dimension a customer may want guarded acceptance; for an incoming inspection screen, simple acceptance may be acceptable.


Step 1: agree the rule before work starts


Clause 7.1.3 applies at contract review. When the customer asks for a statement of conformity to a specification or standard, the specification or standard and the decision rule must be clearly defined. Unless the rule is inherent in the requested specification or standard, the lab tells the customer which rule it will use and agrees it with them.


So the decision rule belongs in the quote or the contract, not in the report. A request, tender and contract review form should ask three questions: is a statement of conformity requested, to which specification, and which decision rule applies?


Some product standards and regulations already define how to treat uncertainty. In that case the rule is inherent in the specification and the lab applies it.


Step 2: document the rule and the risk


When the lab gives a statement of conformity, clause 7.8.6.1 requires it to document the decision rule it uses, taking into account the level of risk associated with it. The standard names three aspects: false accept, false reject and statistical assumptions. The lab then applies the rule.


If the rule is set by the customer, by regulations or by normative documents, no further consideration of risk is needed. If the lab chose or proposed the rule, its record should show the risk it accepts. A decision rule record should capture the specification and limits, the measurement uncertainty, the rule type and any guard band, the resulting acceptance limits, the risk considered, and evidence of the customer's agreement.


Step 3: report it clearly


Clause 7.8.6.2 sets out what the report must make clear about the statement of conformity:


  • which results the statement applies to;

  • which specifications, standards or parts of them are met or not met;

  • the decision rule applied, unless it is inherent in the requested specification or standard.


When the result is close to a limit, measurement uncertainty usually has to be in the report as well. For test reports, uncertainty is reported in the same unit as the measurand, or in a term relative to it, when it is relevant to the validity or application of the results, when the customer asks for it, or when it affects conformity to a specification limit (clause 7.8.3.1). Calibration certificates always include the measurement uncertainty (clause 7.8.4.1).


A worked example


A calibration lab is asked to state whether a customer's pressure gauge is within its ±0.5% of full-scale accuracy class. At contract review the customer agrees to guarded acceptance with a guard band equal to the lab's expanded uncertainty of 0.1% of full scale. The acceptance limit at each point is therefore ±0.4% of full scale.


At one test point the error is 0.45% of full scale. That is inside the specification but outside the acceptance limit, so the certificate states that the gauge does not conform at that point under the agreed guarded acceptance rule. It lists the points covered, the accuracy class used as the specification, the decision rule, and the measured errors with their uncertainty.


Statements of conformity versus opinions


A statement of conformity compares a result with a specification. An opinion or interpretation goes further, for example a judgment on whether an item is fit for a particular use. Clause 7.8.7 allows opinions and interpretations only from authorized personnel, requires the basis to be documented, and requires them to be clearly identified as opinions and interpretations in the report. When they are given in a dialogue with the customer, a record of that dialogue is kept.


Clause 6.2.6 lists analysing results, including statements of conformity and opinions and interpretations, among the activities that need specific authorization. Name the people authorized to sign each kind of statement.


Common gaps


  1. Pass/fail with no rule. Reports state compliance, but no decision rule was defined or agreed.

  2. Rule chosen after the result. The rule is picked when a result is borderline, which undermines impartiality.

  3. Rule missing from the report. The decision rule applied is not stated, and it is not inherent in the specification.

  4. Uncertainty not evaluated. The lab cannot apply a guard band because it never evaluated measurement uncertainty for the method.

  5. Opinions not identified. Comments on fitness for use appear in the report without being marked as opinions and interpretations.


Putting it in place


The ISO/IEC 17025 Laboratory Quality System covers decision rules from contract review to the report. The Review of Requests, Tenders and Contracts Procedure (SOP 7-101-LAB) and its review form (FORM 0115-LAB) capture the specification and the agreed rule. The Reporting of Results Procedure (SOP 7-109-LAB) applies the rule and controls opinions and interpretations. The Decision Rule Record (FORM 0124-LAB) documents the rule, the risk and the customer's agreement. The Measurement Uncertainty Budget workbook (FORM 0121-LAB) gives you the uncertainty that the rule depends on.



Frequently asked questions


Is a decision rule required for every test report?


No. It is required when the lab gives a statement of conformity to a specification or standard. A report that gives results without a pass/fail statement does not need one.


Does ISO/IEC 17025 require guard bands?


No. The standard requires the lab to define, agree, document and apply a decision rule that takes risk into account. Guard bands are one way to do that; simple acceptance can also be agreed when the customer accepts the risk.


What if the specification already says how to treat uncertainty?


Then the decision rule is inherent in the specification. The lab applies it, does not need to agree it separately with the customer, and does not need to name it in the report.