Calibration and Metrological Traceability Under ISO/IEC 17025
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 24, 2026.
Equipment control is where many laboratory problems start. A balance with an overdue calibration, a thermometer with no certificate, or a reference standard nobody can trace back to the SI is easy to find in an audit and hard to explain.
This guide explains what ISO/IEC 17025:2017 requires for equipment, calibration and metrological traceability (clauses 6.4 and 6.5, with Annex A), and how to build a calibration program that meets it. Each requirement is summarized in our own words.
What counts as equipment
The standard takes a broad view. The lab must have access to the equipment it needs for correct work that can influence results, and that includes measuring instruments, software, measurement standards, reference materials, reference data, reagents, consumables and auxiliary apparatus (clause 6.4.1). Equipment outside the lab's permanent control, such as a customer's instrument used on site or rented equipment, must meet the same requirements (clause 6.4.2).
So the equipment inventory is not only a list of instruments. It should include the software that calculates results, the reference materials you calibrate against and the reference data you rely on.
Handling, verification and fitness for purpose
The lab needs a procedure for handling, transport, storage, use and planned maintenance of equipment, so that it works properly and is not contaminated or deteriorated (clause 6.4.3). Before equipment is placed into service, or returned to service after repair or loan, the lab verifies that it meets specified requirements (clause 6.4.4).
Clause 6.4.5 adds a requirement that is easy to miss: measuring equipment must be capable of the accuracy or measurement uncertainty needed for a valid result. A calibration certificate proves that an instrument was checked, not that it is good enough for your method. Compare the instrument's calibrated uncertainty with what the method or the customer's tolerance needs before you put it to use.
When calibration is required
Measuring equipment must be calibrated when its accuracy or measurement uncertainty affects the validity of reported results, or when calibration is needed to establish the metrological traceability of results (clause 6.4.6). Equipment that meets neither test, such as a timer used only to pace a step that has wide limits, may not need calibration, but record that decision.
The lab sets up a calibration program and reviews and adjusts it as needed to keep confidence in calibration status (clause 6.4.7). That review is where intervals change: shorten an interval when an instrument is often found out of tolerance, and lengthen it, with evidence, when results are stable over several cycles.
Equipment that needs calibration, or has a defined validity period, is labelled, coded or otherwise identified so users can see its status (clause 6.4.8). A label with the calibration date and next due date, or an equipment ID that links to the schedule, both work.
Metrological traceability
Metrological traceability means the result can be related to a reference through a documented, unbroken chain of calibrations, each of which adds to the measurement uncertainty (clause 6.5.1). For most labs that reference is the International System of Units (SI).
Clause 6.5.2 gives three routes to SI traceability:
Calibration by a competent laboratory. A calibration lab that meets ISO/IEC 17025 is considered competent.
Certified values of certified reference materials from a competent producer with stated traceability to the SI. A producer meeting ISO 17034 is considered competent.
Direct realization of SI units by comparison, directly or indirectly, with national or international standards.
When traceability to the SI is not technically possible, the lab shows traceability to another appropriate reference, such as certified values of certified reference materials from a competent producer, or results of reference measurement procedures, specified methods or consensus standards (clause 6.5.3).
In practice, the traceability evidence for each instrument is its calibration certificate. Check every incoming certificate before you accept it: it should come from a lab whose competence you have evaluated, state the measurement uncertainty, and say how the results are traceable. ISO/IEC 17025 requires every calibration certificate to state the measurement uncertainty and how the measurements are metrologically traceable (clause 7.8.4), and the provider evaluation comes from clause 6.6.
Intermediate checks and correction factors
Between calibrations, the lab performs intermediate checks when they are needed to keep confidence in equipment performance, and follows a procedure for them (clause 6.4.10). Typical examples are a daily check weight on a balance, an ice point check on a thermometer, or a check standard measured at the start of each run.
When calibration or reference material data include reference values or correction factors, the lab updates them and puts them into use as needed (clause 6.4.11). If a new certificate gives a different correction for a thermometer, the spreadsheet or LIMS that applies the correction must be updated. The lab also takes practicable measures to stop unintended adjustments from invalidating results (clause 6.4.12), for example by sealing adjustment points or locking software settings.
Equipment out of service
Equipment that has been overloaded or mishandled, gives questionable results, or is found defective or outside specified requirements is taken out of service. It is isolated, or clearly labelled or marked as out of service, until it has been verified to work correctly (clause 6.4.9).
The lab then examines the effect of the defect or deviation on earlier work and starts its nonconforming work procedure (clause 7.10). This is the step that is most often skipped. If a reference thermometer is found 0.8 °C out on calibration, every result produced with it since the last good calibration has to be assessed, and customers notified where results may be affected.
Equipment records
Clause 6.4.13 lists the records to keep for each item of equipment that can influence lab activities, where applicable:
identity of the equipment, including software and firmware version;
manufacturer, type and serial number or other unique identification;
evidence of verification that it meets specified requirements;
current location;
calibration dates, results, adjustments, acceptance criteria, and the next due date or calibration interval;
documentation of reference materials, results, acceptance criteria, relevant dates and validity period;
the maintenance plan and maintenance carried out, where relevant to performance;
details of any damage, malfunction, modification or repair.
Most labs keep this as an equipment inventory and calibration schedule for the whole lab, plus an equipment history record for each item.
A calibration program in six steps
Inventory. List every item that can influence results, including software and reference materials, with a unique ID.
Decide. For each item, decide whether calibration is required and record why.
Specify. Set calibration points, ranges and the uncertainty you need, based on the methods that use the equipment.
Source. Choose calibration providers that are competent and can give traceable results with the uncertainty you need, and evaluate them as external providers (clause 6.6).
Schedule and check. Set intervals, schedule intermediate checks, and review every certificate on receipt against your acceptance criteria.
React. When equipment fails a check or calibration, take it out of service and assess its effect on past results.
Common gaps
Certificates accepted without review. Nobody checks that the uncertainty and the results meet the lab's needs.
Out-of-tolerance found, past results ignored. The instrument is adjusted and returned to service with no impact assessment.
Reference materials missing from the inventory. They are part of the traceability chain and need the same control.
Software and firmware versions not recorded. Clause 6.4.13 asks for them.
Intervals never reviewed. Clause 6.4.7 expects the calibration program to be reviewed and adjusted.
Putting it in place
The ISO/IEC 17025 Laboratory Quality System includes the Equipment Management Procedure (SOP 7-103-LAB) and the Calibration Program and Metrological Traceability Procedure (SOP 7-104-LAB). It also includes the Equipment Inventory and Calibration Schedule workbook (FORM 0109-LAB), which calculates each next due date and flags past-due items, plus the Equipment History Record, Intermediate Check Record, Out-of-Service Equipment and Impact Assessment, and Reference Material Register.
Related reading: ISO/IEC 17025 Requirements: A Practical Guide for Testing and Calibration Labs and Decision Rules and Statements of Conformity Under ISO/IEC 17025.
Frequently asked questions
Does every piece of lab equipment need calibration under ISO/IEC 17025?
No. Calibration is required when the equipment's accuracy or measurement uncertainty affects the validity of reported results, or when it is needed for metrological traceability. Record the decision for equipment that is not calibrated.
Can a lab calibrate its own equipment?
Yes, if it can show traceability, for example through reference standards that were themselves calibrated by a competent lab. A lab that calibrates its own equipment must evaluate measurement uncertainty for those calibrations (clause 7.6.2).
Should a calibration certificate state the next due date?
The lab sets its own intervals. ISO/IEC 17025 says a calibration certificate or label must not contain a recommended calibration interval unless this has been agreed with the customer (clause 7.8.4).