Root Cause Analysis for CAPA: 5 Whys, Fishbone, Is / Is Not and Proving the Cause
Reviewed by Aaron Lyon, President of A.P. Lyon and former Director of Quality Assurance and International Regulatory, with more than 30 years in medical device quality. Last reviewed November 28, 2026.
A CAPA is only as good as its root cause. If the cause is wrong, every action that follows is aimed at the wrong target, the effectiveness check fails or, worse, passes by luck, and the problem comes back months later under a new CAPA number. Most weak investigations fail in the same few places: a vague problem statement, a cause chosen before the data is in, "operator error" accepted as an answer, and no test of whether the cause actually explains what happened.
This guide sets out a practical route for root cause analysis within a CAPA, for ISO 13485, the FDA QMSR and ISO 9001:2026. The requirements are summarized in our own words; check the standards and the official text of 21 CFR 820 before relying on them.
What the standards require, and what they leave to you
ISO 13485:2016 clause 8.5.2 b) requires the CAPA procedure to define requirements for determining the causes of nonconformities, and the clause requires records of the results of any investigation. Clause 8.5.3 a) asks for potential nonconformities and their causes to be determined. ISO 9001:2026 clause 10.2.1 b) asks the organization to review the nonconformity, determine its causes and determine whether similar nonconformities exist or could potentially occur, and clause 10.2.2 requires documented information on the nature of the nonconformities and the actions taken. In the United States, the QMSR applies ISO 13485 as a requirement through 21 CFR 820.10(a).
None of these documents names a method. 5 Whys, fishbone diagrams, is / is not analysis, fault trees and Pareto charts are widely used good practice, but no standard requires any of them. What is required is that the cause is determined and the investigation is recorded. That leaves you free to choose the method that fits the problem, as long as the record shows how you got to the cause and why you believe it.
Step 1: plan the investigation
Before collecting anything, write down who is on the team, what is in scope (products, lots, processes, sites), what data you need and when the investigation is due. The team should include the CAPA owner, at least one person who works where the problem occurred, and someone from engineering or operations who knows the process. Investigation depth should follow the risk: a low-priority CAPA may need a clear problem statement, a 5 Whys and its evidence; a high-priority CAPA should use at least two methods and prove the cause by test or data.
For a product failure, a returned product or a complaint, plan the failure analysis and keep the returned product and retained samples on hold until the investigation is closed. Evidence that has been scrapped cannot be re-examined when the first theory turns out to be wrong.
Step 2: write a problem statement that holds only facts
The problem statement is the foundation. It answers what (the object and the defect, and the requirement that was not met), where, when, who found it, how it was detected, and how many. It contains numbers and facts, and it leaves out causes and solutions.
Weak: "Pouch seals are bad because the night shift is not careful."
Strong: "14 of 2,400 pouches from lot 2611-07 failed the seal peel test (requirement 1.2 to 2.8 N/15 mm) at final inspection on line 2, night shift, 26 November."
The weak version has already chosen a cause and a culprit. The strong version gives the team something to test every theory against.
Step 3: draw the is / is not picture
For each line of the problem statement, record where the problem is found and where it could be found but is not. If it is on line 2 but not line 1, on the night shift but not days, at the far edge of the seal but not the near edge, after four hours of running but not at the start, write each pair down. Then list what is different or what changed between the two columns.
Is / is not is the most underused tool in CAPA work. It costs half an hour, and it rules out a large share of candidate causes before anyone builds a fishbone. Any true root cause must explain both columns: why the problem is where it is, and why it is not where it is not.
Step 4: collect data and check the extent
Gather the evidence: production and inspection records, nonconformance reports, complaints, service reports, process data, calibration and maintenance records, supplier records, change history and training records. Use statistical methods where they help show a trend or a difference. ISO 13485 clause 8.4 requires procedures for analysis of data that include determining appropriate methods, including statistical techniques.
Then check the extent. Could the same problem, or the same cause, exist elsewhere: other lots, products, lines, sites, suppliers or markets? ISO 9001:2026 clause 10.2.1 b) 3) asks this directly. Record what you checked and the result. Where the cause could act but has not yet, raise a preventive action on the same CAPA. If the extent check finds more affected product, or product already shipped, go back to containment at once.
Step 5: choose the method that fits the problem
5 Whys: for one clear failure with a straight chain of cause and effect. Ask why and answer with a fact, then ask why again of that answer, until you reach a cause the company controls and could remove.
Fishbone (cause and effect): for many possible causes, or a team that disagrees. List candidate causes under Method, Machine, Material, Measurement, People and Environment (plus Design for devices), then test each against the evidence.
Is / is not: for a problem that appears in some places and not others. Compare where it is and is not; the differences are the clues.
Fault tree: for serious or complex failures with several conditions acting together. Break the top event into the events that must occur, joined by AND and OR gates.
Pareto chart: for many events of several types. Count by type, sort largest first, and work on the few types that make up most of the total.
The table is a guide, not a rule. Use a second method when the first does not produce a cause that the evidence supports.
Step 6: run two chains, not one
Most problems have two causes worth finding: why it happened (occurrence) and why it was not caught earlier (detection). Run a 5 Whys for each. In the pouch example, the occurrence chain might end at "the sealer platen temperature drifts low after four hours and nothing checks it during long runs"; the detection chain at "peel samples are taken only at the start of the run". Fixing only the first leaves you blind if a different cause produces the same defect; fixing only the second catches bad product but keeps making it.
Step 7: do not stop at human error
"Operator error", "did not follow the procedure" and "needs retraining" appear in many CAPA files as the root cause. They are almost never the end of the chain. Ask why the system allowed the error: was the instruction unclear, was there no error-proofing, was the workload too high, was the layout confusing, was there no check? Record the system cause. Retraining may still be one of the actions, but if it is the only action, expect the problem to return when the next person joins.
Step 8: prove the cause
A cause is a hypothesis until it has been tested against the evidence. For each candidate, ask whether it explains every "is" and every "is not". Where you can, prove it: show it in the data, switch it on and off and watch the defect follow, or reproduce the failure. Record the test and its result.
Then record the conclusion honestly. "Confirmed" means you proved it by test or data. "Probable" means it is the best supported cause but could not be proved, with the reason. A probable cause is acceptable, but it needs extra monitoring during the effectiveness check, because there is a real chance the problem returns for a reason you did not find.
Step 9: write the summary and get it reviewed
The investigation summary states the problem, the extent, the root cause and any contributing causes, the confidence and the evidence references. The quality manager reviews it, and returns it with comments if the cause is not supported. For a device complaint, the complaint record is updated with the investigation result and any correction or corrective action; under the QMSR, 21 CFR 820.35(a) lists the information complaint records must hold for investigated complaints, including any correction or corrective action taken.
Only then do actions get planned, and their success measured. That part is covered in CAPA Effectiveness Checks: Setting Criteria Before You Act and Closing With Evidence.
Common root cause mistakes
Starting with the answer. The team "knows" it was the new supplier and collects only the data that agrees.
A problem statement with a cause in it. It closes off every other line of inquiry before the work starts.
Stopping at a symptom. "The seal was weak" is a description, not a cause.
One chain only. The detection cause goes unaddressed and the next escape is just as likely.
Human error as the root cause. The system that allowed the error is untouched.
No extent check. The same cause is waiting on the sister product.
No test of the cause. The effectiveness check becomes the first real test, months later.
Putting it into practice
The CAPA and Root Cause Kit turns this route into numbered steps. SOP 8-102-CAP covers planning, the problem statement, is / is not, data and extent, choosing and using the methods, and proving the cause; FORM 0103-CAP gives you the problem statement, is / is not table, two-chain 5 Whys, fishbone, cause verification and summary on one record; and the Excel CAPA log tracks the root cause category so the review board can see which kinds of cause keep recurring. It comes in ISO 13485, dual-standard and ISO 9001 editions. For deciding which problems need this depth of investigation, see When Does a Problem Need a CAPA? Triage, Containment and Risk Rating.
Frequently asked questions
Does ISO 13485 require the 5 Whys or a fishbone diagram?
No. The standard requires the causes of nonconformities to be determined and the investigation to be recorded. The methods are tools for doing that well; choose the ones that fit the problem and record how you used them.
What if we cannot find the root cause?
Record the most probable cause with the evidence for it and the reason it could not be confirmed, put interim controls in place, and add monitoring during the effectiveness check that would show the problem returning. If it does return, the CAPA is reopened and the investigation goes further.
How many whys are enough?
Usually three to six. Stop when you reach a cause the company controls and could remove, and which, if removed, would have prevented the problem. Stopping earlier leaves a symptom; going further tends to reach causes outside your control.
Building a complete quality system? Build Your QMS lets you choose ISO 13485 + FDA QMSR, Dual-Standard or ISO 9001:2026 procedures in one place, and applies volume discounts automatically: 20% off 5–9 procedures, 35% off 10–19, 50% off 20–39 and 60% off 40 or more.
Not sure where your quality system stands? Take the free QMS gap check: about 20 questions, an instant report of your gaps by clause, and the procedures that close each one.
Just getting started? The free Startup Quality Plan shows which procedures you need now, which can wait, and how long they take to put in place.



Comments